Shimadzu Xdimensus 300
Shimadzu ·Made in Japan · In Active Production

Shimadzu Xdimensus 300

Shimadzu Xdimensus 300 — Spectrometers.

The Shimadzu Xdimensus 300 is a dimensional X-ray CT system with a 225 kV, 135 W generator and flat-panel detector, configured for non-destructive internal and external metrology of resin and light-metal parts up to ø300 mm × 210 mm and 10 kg. It achieves a sphere-distance error of ±(3.8 + L/50) µm and probing error PS/PF = 5.0 µm, making it suited to first-article inspection and dimensional verification where internal features are inaccessible.

Full specifications

Core Performance & Functional Specifications
Sphere distance measurement error (XDimensus 300; example 5.8 μm for L = 100 mm) mfr (3.8 + L/50)* μm (XDimensus 300; example 5.8 μm for L = 100 mm)
Measurable sample diameter mfr ø300 mm (XDimensus 300)
Dimensions, Weight & Mechanical Design
Measurable sample height mfr H 210 mm (XDimensus 300)
Cabinet width mfr 2,195 mm (XDimensus 300)
Cabinet depth mfr 1,561 mm (XDimensus 300)
Cabinet height mfr 1,971 mm (XDimensus 300)
Installation space width mfr 4000 mm (XDimensus 300)
Installation space depth mfr 3500 mm (XDimensus 300)
Electrical, Power & Drive Specifications
Max. tube voltage mfr 225 kV (XDimensus 300)

Description

Construction and positioning

Integration & Operating Notes

Electrical
Three separate power feeds with different phases and voltages

The main unit requires single-phase 200 V (4 kVA), the air-conditioning unit requires three-phase 200 V (2.6 kVA), and the control computer requires single-phase 100 V (1.5 kVA). Each must be supplied from an appropriate branch circuit; sharing a feed or omitting the three-phase supply will prevent normal operation.

Factory Specification Sheet
Process Limits
Samples limited to 300 mm diameter, 210 mm scan height, and 10 kg

The maximum CT scan region is ø300 mm × H 210 mm, and the maximum sample size and weight are 300 mm and 10.0 kg. Exceeding these limits risks collision with the source-detector geometry or inaccurate reconstruction; fixturing must keep the sample within the scan envelope.

Factory Specification Sheet
Cooling & Media
Compressed air supply of 0.6–1 MPa at 60 SLPM minimum required

The system requires an air source delivering 0.6 MPa to 1 MPa at a minimum flow of 60 SLPM. This air is used by the air-conditioning unit and possibly for pneumatic functions; a stable, dry supply must be available at the installation site.

Factory Specification Sheet
Electrical
Type-D ground with 100 Ω max resistance is mandatory

A dedicated Type-D ground with ground resistance not exceeding 100 Ω is specified. This is required for electrical safety and to minimize noise in the low-current detector electronics; the ground must be verified before commissioning.

Factory Specification Sheet

Selection & Integration Q&A

What are the power and air supply requirements for integrating this CT system into my facility?

Provide single-phase 200 V 4 kVA for the main unit, three-phase 200 V 2.6 kVA for the air-conditioning unit, and single-phase 100 V 1.5 kVA for the control computer, plus a Type-D ground (100 Ω max) and air at 0.6–1 MPa, 60 SLPM min.

What floor loading and environmental conditions are required for installation?

The shield box weighs approx. 4,000 kg and the air-conditioning unit approx. 200 kg; the floor must support these loads. External leakage dose is 1 µSv/h max, so radiation shielding is integral to the cabinet and no additional enclosure is specified.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c251-e035
Datasheet ·3,0 МБ ·12 pp.
View PDF