No photo available yet

Waters ·Made in USA · In Active Production

Waters 2475 Fluorescence (FLR) Detector for HPLC Systems

Waters 2475 Fluorescence (FLR) Detector for HPLC Systems — Chromatography detectors.

The Waters 2475 Fluorescence (FLR) Detector is a xenon-lamp-based HPLC detector with axially illuminated 13 µL flow cell, covering 200–890 nm excitation and 210–900 nm emission. It provides high sensitivity (Raman S/N ≥1000) and up to four 2D channels or one 3D channel, serving as a selective fluorescence detector in chromatographic systems.

Full specifications

Core Performance & Functional Specifications
Detector type Fluorescence (FLR)
Excitation wavelength range mfr 200 to 890 nm
Emission wavelength range mfr 210 to 900 nm
Bandwidth mfr 20 nm
Wavelength accuracy mfr ±3 nm
Wavelength repeatability mfr ±0.25 nm
Sensitivity mfr S/N, Raman peak of water ≥1000
Measurement range mfr 0.001 to 100,000.000 emission units
Filter time constant, single λ mode mfr 0.1 to 5.0 s Hamming (default); 0.1 to 99.0 s (o
Filter time constant, dual λ mode mfr 1.0 to 50.0 s Hamming (default); 1.0 to 99.0 s (
Data channels mfr Up to four 2D channels or one 3D channel
Sampling rate mfr Up to 20 points/s in single mode; 1 point/s in 3
Light source mfr Xenon lamp, 150 W
Flow cell design mfr Axially illuminated
Flow cell volume mfr 13 μL
Pressure limit mfr 145 psi (10 bar)
Wetted materials mfr 316 stainless-steel, fused silica, Teflon
Power requirements mfr 100 to 240 VAC
Inputs mfr Four event inputs
Outputs mfr Four outputs (2 analog, 2 event)
Dimensions, Weight & Mechanical Design
Weight 18.1 kg
Electrical, Power & Drive Specifications
Power consumption 280 VA
Line frequency mfr 50 to 60 Hz

Description

Construction and positioning

Integration & Operating Notes

Electrical
Ensure supply can deliver 280 VA at 100–240 VAC

The detector draws 280 VA; verify the branch circuit and any UPS can handle this continuous load plus inrush. Operation outside 100–240 VAC or 50–60 Hz may cause malfunction or damage.

Factory Specification Sheet
Process Limits
Do not exceed 145 psi (10 bar) in the flow cell

The 13 µL flow cell is rated to 145 psi (10 bar). Exceeding this pressure can rupture the cell or cause leaks, invalidating results and risking operator safety. Install a backpressure regulator or pressure relief upstream if system pressures may spike.

Factory Specification Sheet
Maintenance
Replace xenon lamp when sensitivity drops below specification

The 150 W xenon lamp has finite life; when Raman S/N of water falls below 1000, lamp replacement is indicated. Follow manufacturer's procedure for lamp replacement and recalibration of wavelength accuracy to ±3 nm.

Factory Specification Sheet

Selection & Integration Q&A

What are the power and I/O requirements for integrating the 2475 into an HPLC system?

The detector requires 100–240 VAC, 50–60 Hz, 280 VA. It provides four event inputs and four outputs (2 analog, 2 event) for synchronization with pumps, autosamplers, and data systems. Ensure the controller or software supports event-based triggering and analog signal acquisition.

What maintenance items and accessories are required but not included in the base delivery?

The base delivery does not include cables for event/analog connections, replacement xenon lamps, or flow cell spare parts. A maintenance schedule should include lamp replacement when S/N drops below 1000 and periodic cleaning of the flow cell. Refer to the manual for specific intervals.