Waters 3465 Electrochemical Detector for LC Systems
Waters ·Made in USA · In Active Production

Waters 3465 Electrochemical Detector for LC Systems

Waters 3465 Electrochemical Detector for LC Systems — Chromatography detectors.

The 3465 Electrochemical Detector is a single- or dual-flow-cell LC detector supporting DC, PULSE, PULSE 2, and SCAN modes with a 10 pA–200 µA current range and integrated column/flow-cell heating to 60 °C. It delivers <2 pA noise in the 1 nA range and ±2.5 V analog output, serving as a sensitive, temperature-controlled electrochemical detector in HPLC systems.

Full specifications

Core Performance & Functional Specifications
Detector type Electrochemical Detector
Operating modes mfr Direct Current (DC), DC, PULSE, PULSE 2, SCAN
Other modes mfr CONFIG, DIAG, and SERVICE
Detector configurations mfr Single Flow Cell (SCC) (p/n: 725000673) or Dual
Offset mfr +50% to -50% of max. output voltage, 5% steps
Analog signal output (DAC) mfr -1 to +1 V full scale (via 16-bit D/A converter)
Analog output (I/E) mfr -2.5 to +2.5 V full scale (unprocessed I/E conve
Integrated flow cell and column heating compartment mfr +7 °C above ambient to 60 °C, accuracy 0.5 °C, s
DC Mode Range mfr 10 pA–200 μA in 1, 2, or 5 increments
DC Mode Filter (ADF) mfr 10–0.001 Hz in 1, 2, or 5 increments; RAW and OF
DC Mode Potential (Ec) mfr -2.5 V to +2.5 V with 10 mV increments
DC Mode Data rate mfr 1–100 Hz in 1, 2, or 5 increments, dependent on
Pulse Mode Range mfr 10 nA–200 μA in 1, 2, or 5 increments
Pulse Mode Filter (ADF) mfr 0.5–0.001 Hz in 1, 2, or 5 increments; OFF: For
Pulse Mode Potential (Ec) mfr -2.5 V to +2.5 V with 10 mV increments
Pulse Mode Data rate mfr 1/(pulse duration) Hz
Pulse Mode Waveform mfr Max five potential steps
Pulse times (t1-t2) mfr t1: 100 ms–2000 ms; t2, t3, t4, and t5: 0–2000 m
Sampling times (ts) mfr 20 ms – [t1–60] ms
Pulse Mode 2 Range mfr 10 nA–200 μA in 1, 2, or 5 increments
Operating system mfr Windows 10
Dimensions, Weight & Mechanical Design
Weight 14.4 Kg
Height mfr 44.0 cm
Width mfr 22.0 cm
Depth mfr 43.0 cm
Height mfr 44.0 cm (17.3 inches) (packed instrument (shipping box))
Width mfr 22.0 cm (8.7 inches) (packed instrument (shipping box))
Depth mfr 43.0 cm (16.9 inches) (packed instrument (shipping box))
Electrical, Power & Drive Specifications
Power consumption 260 VA
Max. current compensation (Autozero) mfr 25 nA - 2.5 mA in DC and PULSE mode dependent on
Operating Conditions, Environment & Protection
Noise <2 pA with dummy cell (load of 300 MΩ/470 pF) in 1 nA range, filter off, Ec +800 mV, and temp. of 35 °C
Storage humidity range mfr 20% to 80%, non-condensing %

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Flow cell selection must match single or dual configuration

The detector accepts either a single flow cell (p/n 725000673) or dual flow cell (p/n 725000686). Using an incompatible cell or mixing components can prevent proper sealing and electrical contact, risking leaks or invalid electrochemical response.

Factory Specification Sheet
Control & I/O
Analog outputs require compatible data acquisition

The DAC output is −1 to +1 V full scale via 16-bit converter; the unprocessed I/E signal is −2.5 to +2.5 V full scale. Ensure the connected integrator or A/D system accepts these ranges and has sufficient resolution to capture the 16-bit DAC signal without clipping.

Factory Specification Sheet
Process Limits
Pulse mode timing must satisfy sampling window

In pulse modes, sampling time (ts) must be between 20 ms and (t1−60) ms. Setting t1 below 80 ms or ts outside this window will produce invalid sampling and unreliable peak data; verify waveform timing before acquisition.

Factory Specification Sheet

Selection & Integration Q&A

What are the requirements for connecting the 3465 to my LC system and data acquisition?

The 3465 provides a 16-bit DAC output of −1 to +1 V full scale and an unprocessed I/E output of −2.5 to +2.5 V full scale. Ensure your integrator or A/D accepts these ranges. The flow cell must be either SCC (725000673) or DCC (725000686).

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Waters 3465 Electrochemical Detector Overview and Maintenance Guide
Brochure ·6,4 МБ ·192 pp.