No photo available yet

Waters ·Made in USA · In Active Production

Waters ACQUITY Arc System

Waters ACQUITY Arc System — Chromatographs with Flow rate 5 milliliter/minute, Max pressure 65.5 megapascal.

The ACQUITY Arc System is a quaternary, stainless-steel flow-path chromatograph with a 5 mL/min maximum flow rate and 9500 psi (65.5 MPa) maximum pressure, housed in a 711 cm high, 343 cm wide cabinet. It supports column temperatures to 90 °C, sample compartment control from 4–40 °C, and up to 15 columns of 150 mm length. It serves as a versatile LC platform for method transfer and routine analysis across multiple detector types.

Flow rate
5 milliliter/minute
Max pressure
65.5 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 5 mL/min
Max pressure 9500 psi
Column temperature 90 °C
Sample flow path mfr Stainless-steel
Sample carryover mfr ≤0.002% (Caffeine) under UV conditions
Sample compartment mfr 4–40 °C
Pump options mfr Quaternary
Solvent conditioning mfr Passive and active pre-heating
Column capacity mfr Up to 15 columns of 150 mm length and up to 4.6
Maximum sample capacity mfr Up to 20 plates (with Sample Organizer)
Detector options mfr TUV Detector, PDA Detector, FLR Detector, Charge
External control mfr Empower Software, MassLynx Software
Dimensions, Weight & Mechanical Design
Height 711 cm
Width 343 cm
Depth mfr 71.2 cm (28.0 in.)
Materials, Media & Process Compatibility
pH range mfr 1–12.5

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Mobile phase pH must stay within 1–12.5

The stainless-steel flow path tolerates pH 1–12.5. Outside this range, corrosion and sample adsorption may occur, invalidating results and shortening component life. Verify buffer pH before use, especially with aggressive modifiers.

Factory Specification Sheet
Control & I/O
Requires Empower or MassLynx for external control

The system is controlled via Empower or MassLynx software. Ensure a compatible workstation with the correct software version and appropriate network or local connection is available before installation; no standalone keypad control is described in the supplied facts.

Factory Specification Sheet

Selection & Integration Q&A

What detector and software compatibility does the ACQUITY Arc System offer?

It supports TUV, PDA, FLR, Charged Aerosol, RI, ELS, and EC detectors, and is controlled via Empower or MassLynx software. Confirm the specific detector model and software version with Waters before ordering, as not all combinations may be supported.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation