Waters ACQUITY Premier System with Multi-Dimensional Technology
Waters ·Made in USA · In Active Production

Waters ACQUITY Premier System with Multi-Dimensional Technology

Waters ACQUITY Premier System with Multi-Dimensional Technology — Chromatographs with Flow rate 2 milliliter/minute, Max pressure 103.4 megapascal.

The ACQUITY Premier System with Multi-Dimensional Technology is a high-pressure liquid chromatograph configured for automated multi-dimensional separations, integrating binary or quaternary solvent delivery with a 15,000 psi-capable pump and dual column compartments. It supports flow rates from 0.001 to 2.000 mL/min and injection volumes from 0.1 to 10.0 µL as standard, with optional extension loops up to 1000 µL. The system is intended for method development and complex sample analysis requiring orthogonal separation dimensions within a single instrument footprint.

Flow rate
2 milliliter/minute
Max pressure
103.4 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 2.0 mL/min
Max pressure 15000 psi
Column temperature 90.0 °C
Injection volume 10.0 μL
Number of solvents mfr Up to four, in any combination of two: Al or A2
Solvent conditioning mfr Five vacuum degasser, one allocated for injector
Operating flow rate range mfr 0.001 to 2.000 mL/min, in 0.001 mL increments (f
Maximum operating pressure mfr 15,000 psi up to 1 mL/min, 9000 psi up to 2 mL/m
Injection volume range mfr 0.1 to 10.0 μL as standard; Up to 1000.0 μL with
Sample compartment temp. mfr 4.0 to 40.0 °C, settable in 0.1 °C increment
Column compartment(s) temperature range mfr 4.0 to 90.0 °C, settable in 0.1 °C increments
Operating temperature range mfr 4.0 to 40.0 °C (39.2 to 104.0 °F)
Dimensions, Weight & Mechanical Design
Height 1034 mm
Width 838 mm
Depth mfr 86.4 cm (34 in.)
Electrical, Power & Drive Specifications
Voltage mfr 100 to 240 VAC
Frequency mfr 50 to 60 Hz
Operating Conditions, Environment & Protection
Operating humidity range mfr 20% to 80%, non-condensing
Acoustic noise mfr <62 dBA

Description

Construction and Architecture

Integration & Operating Notes

Process Limits
Do not exceed 9000 psi above 1 mL/min per pump

The pump's maximum pressure is 15,000 psi only up to 1 mL/min; between 1 and 2 mL/min the limit drops to 9000 psi per pump, and the combined system pressure must never exceed 15,000 psi. Exceeding these limits risks seal failure and invalid results.

Factory Specification Sheet
Environment
Ambient must stay within 4–40 °C and 20–80% RH non-condensing

Operation outside this range can affect detector baseline stability, column thermostat performance, and electronics reliability. Ensure laboratory HVAC maintains these conditions continuously, especially when the system is used unattended overnight.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this multi-dimensional ACQUITY Premier system instead of a standard single-dimension ACQUITY Premier?

Choose this model when your method requires automated orthogonal separations—e.g., heart-cutting or comprehensive 2D-LC—within one instrument. A standard single-dimension system suffices for routine assays without multi-dimensional capability, and it avoids the added complexity of dual-column switching.

What environmental and duty-cycle limits apply when running the system continuously?

Ambient must remain 4–40 °C and 20–80% RH non-condensing. The system is designed for continuous operation, but pump seals and degasser membranes are wear items; no duty-cycle restriction is stated, yet preventive maintenance should be planned based on solvent use and pressure history.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
ACQUITY Premier System with 2D Technology Capabilities Guide
Manual ·1,4 МБ ·67 pp.
View PDF