Waters ACQUITY UPLC H-Class PLUS System
Waters ·Made in USA · In Active Production

Waters ACQUITY UPLC H-Class PLUS System

Waters ACQUITY UPLC H-Class PLUS System — Chromatographs with Flow rate 2.2 milliliter/minute, Max pressure 103.4 megapascal.

The ACQUITY UPLC H-Class PLUS System is a stainless-steel flow-path UPLC chromatograph with binary or quaternary pumping, active pre-heating, and up to 15,000 psi operation at 2.2 mL/min. It accommodates up to six 150 mm columns and 20 sample plates with the Sample Organizer, positioning it for high-throughput method development and routine analysis. Its open detector architecture supports PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometric detection under Empower, MassLynx, waters_connect, or UNIFI control.

Flow rate
2.2 milliliter/minute
Max pressure
103.4 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 2.2 mL/min
Max pressure 15000 psi
Sample flow path mfr Stainless-steel
Sample carryover mfr ≤0.002% (Caffeine) under UV conditions
Sample compartment mfr 4–40 ̊C
Pump options mfr Binary, Quaternary
Solvent conditioning mfr Active Pre-heating
Column capacity mfr Up to 6 columns of 150 mm length and up to 4.6 m
Maximum sample capacity mfr Up to 20 plates (with Sample Organizer)
Detector options mfr PDA Detector, TUV Detector, FLR Detector, Charge
External control mfr Empower Software, MassLynx Software, waters_conn
Dimensions, Weight & Mechanical Design
Height 711 cm
Width 343 cm
Depth mfr 71.2 cm (28.0 in.)
Materials, Media & Process Compatibility
pH range mfr 1–12.5

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 15,000 psi system pressure

Operating above the 15,000 psi maximum risks seal, tubing, and column hardware failure; method gradients and column dimensions must be selected so that backpressure stays below this limit at the intended 2.2 mL/min flow.

Factory Specification Sheet
Process Limits
Mobile-phase pH must stay within 1–12.5

The stainless-steel flow path tolerates pH 1–12.5; outside this range, corrosion and metal leaching can occur, compromising both hardware life and chromatographic results, especially with MS detection.

Factory Specification Sheet
Control & I/O
External data system is required for control

The system is controlled via Empower, MassLynx, waters_connect, or UNIFI; no standalone control is specified. Integration planning must include the data-system workstation, network, and license before the instrument can run methods.

Factory Specification Sheet
Cooling & Media
Sample compartment must be set within 4–40 °C

Sample stability and injection reproducibility depend on holding the sample compartment within its 4–40 °C range; lab ambient conditions and plate residence time should be checked for thermally labile samples.

Factory Specification Sheet
Maintenance
Plan column and sample capacity around bench access

With up to six 150 mm columns and up to 20 plates with the Sample Organizer, routine column changes and plate loading require unobstructed front and side access; layout the bench so these tasks do not disturb detector or MS connections.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this H-Class PLUS System rather than a binary-only or lower-pressure variant?

Choose this model when you need up to 15,000 psi at 2.2 mL/min, pH 1–12.5, and flexibility across binary or quaternary pumping with active pre-heating. A binary-only variant suits fixed high-pressure gradients; this system adds quaternary mixing and broad detector options for method development.

What adjacent system components and data-system requirements must be in place before installation?

The system requires an external control platform (Empower, MassLynx, waters_connect, or UNIFI) plus the selected detector (PDA, TUV, FLR, charged aerosol, RI, ELS, or MS). Confirm bench space for up to six 150 mm columns and, if used, the Sample Organizer for up to 20 plates.

What mounting, environment, and duty-cycle limits apply to this system?

Operate within 15,000 psi maximum pressure and pH 1–12.5; keep the sample compartment at 4–40 °C. The stainless-steel flow path and active pre-heating require stable lab ambient conditions. No PDF manual was supplied, so confirm duty-cycle and ventilation requirements with Waters.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation