Waters ACQUITY UPLC I-Class PLUS System
Waters ·Made in USA · In Active Production

Waters ACQUITY UPLC I-Class PLUS System

Waters ACQUITY UPLC I-Class PLUS System — Chromatographs with Flow rate 2 milliliter/minute, Max pressure 124.1 megapascal.

The ACQUITY UPLC I-Class PLUS System is a binary, stainless-steel flow path UPLC chromatograph with active pre-heating, rated to 18,000 psi (124 MPa) and 2.0 mL/min. It supports columns up to 150 mm × 4.6 mm I.D. and up to 20 sample plates with the Sample Organizer, distinguishing it within the I-Class series by its PLUS configuration. It serves as the high-pressure separation front end for PDA, TUV, FLR, CA, RI, ELS, or mass spectrometric detection in regulated analytical workflows.

Flow rate
2 milliliter/minute
Max pressure
124.1 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 2.0 mL/min
Max pressure 18000 psi
Column temperature 4–40 °C
Sample flow path mfr Stainless-steel
Sample carryover mfr <0.001% (Caffeine) under UV conditions
Pump options mfr Binary
Solvent conditioning mfr Active Pre-heating
Column capacity mfr Up to 2 columns of 150 mm length and up to 4.6 m
Maximum sample capacity mfr Up to 20 plates (with Sample Organizer)
Detector options mfr PDA Detector, TUV Detector, FLR Detector, Charge
External control mfr Empower Software, MassLynx Software, waters_conn
Dimensions, Weight & Mechanical Design
Height 711 mm
Width 343 mm
Depth mfr 71.2 cm (28.0 in.)
Materials, Media & Process Compatibility
pH Range mfr 1–12.5

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 18,000 psi or 2.0 mL/min in combination

The pump and flow path are rated to 18,000 psi (124.1 MPa) and 2.0 mL/min. Operating near both maxima simultaneously raises wear and leak risk; verify column and tubing pressure ratings before method transfer from lower-pressure HPLC.

Factory Specification Sheet
Process Limits
Mobile phase pH must stay within 1–12.5

The stainless-steel flow path tolerates pH 1 to 12.5. Outside this window, corrosion and altered retention can occur; confirm buffer and additive pH before running extended gradients.

Factory Specification Sheet
Control & I/O
Choose control software before commissioning

The system can be driven by Empower, MassLynx, waters_connect, or UNIFI. The selected platform determines network, licensing, and data-integrity configuration, so decide before installation to avoid rework.

Factory Specification Sheet
Maintenance
Carryover performance depends on flow-path upkeep

The <0.001% caffeine carryover figure under UV conditions is a specification, not a guarantee over time. Establish a preventive maintenance interval for seals, needles, and injection components to preserve it.

Factory Specification Sheet

Selection & Integration Q&A

Which detectors and control software can be integrated with this system?

Detector options include PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometers. Control is via Empower, MassLynx, waters_connect, or UNIFI. Confirm the detector's pressure and flow ratings against the 18,000 psi and 2.0 mL/min system limits before connection.