Waters ACQUITY UPLC M-Class System w/ 2D Technology
Waters ·Made in USA · In Active Production

Waters ACQUITY UPLC M-Class System w/ 2D Technology

Waters ACQUITY UPLC M-Class System w/ 2D Technology — Chromatographs with Flow rate 2×10⁻⁷ milliliter/minute, Max pressure 103.4 megapascal.

The ACQUITY UPLC M-Class System with 2D Technology is a binary, non-ferrous flow-path chromatograph configured for two-dimensional separations at pressures up to 15,000 psi (1,034 bar). It accommodates a single column from 75 µm to 4.6 mm ID and up to 250 mm length, with injection volumes down to 0.1 µL and flow rates as low as 0.2 nL/min. It serves as the primary separation engine in micro- to narrow-bore UPLC workflows requiring orthogonal second-dimension analysis.

Flow rate
2×10⁻⁷ milliliter/minute
Max pressure
103.4 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 0.2 nL/min
Max pressure 15000 psi
Injection volume 0.1 µL
Sample flow path mfr Non-ferrous
Sample compartment mfr 4–40 ̊C
Pump options mfr Binary
Column capacity mfr Single column, 75 μm to 4.6 mm internal diameter
Maximum sample capacity mfr Up to 21 plates (with Sample Organizer)
Detector options mfr PDA Detector, TUV Detector, Mass Spectrometers
External control mfr MassLynx Software
Dimensions, Weight & Mechanical Design
Height 697 cm
Width 343 cm
Depth mfr 71.2 cm (28.0 in.)
Materials, Media & Process Compatibility
pH range mfr 2 to 10

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 15,000 psi system pressure

Operating above the rated maximum pressure risks seal failure, column damage, and invalid results. Ensure method gradients and column backpressure calculations stay below this limit, especially with sub-2 µm particles and high flow rates.

Factory Specification Sheet
Process Limits
Mobile phase pH must remain between 2 and 10

The non-ferrous flow path tolerates pH 2–10; outside this range, corrosion or analyte adsorption may occur. Verify buffer and sample pH before analysis, and flush the system after using extreme pH conditions.

Factory Specification Sheet
Control & I/O
System control requires MassLynx software connection

The instrument has no standalone control interface; all methods, sequences, and 2D valve switching are managed through MassLynx. Ensure the controlling workstation is available and configured before startup, and that the software version supports the 2D Technology configuration.

Factory Specification Sheet
Maintenance
Sample Organizer is an optional accessory, not included

Base delivery does not include the Sample Organizer required to reach the maximum sample capacity of 21 plates. Order it separately if high-throughput or long unattended sequences are needed, and account for its additional bench space and connections.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 2D Technology configuration instead of a standard single-dimension M-Class system?

Choose 2D when your application requires orthogonal separation of complex mixtures that cannot be resolved in a single dimension. A standard M-Class system is sufficient for routine micro- to narrow-bore UPLC separations without a second dimension.

What detector and software are required to operate this system?

The system is compatible with PDA, TUV, or mass spectrometers and is controlled externally via MassLynx software. Ensure the detector and MassLynx version support the 2D Technology configuration and the intended flow rates.