Waters ACQUITY UPLC PDA Detector for Spectral Profile Detection
Waters ACQUITY UPLC PDA Detector for Spectral Profile Detection — Chromatography detectors.
The ACQUITY UPLC PDA Detector is a photodiode array detector with a light-guiding UPLC intelligent flow cell (500 nL, 10 mm path) for spectral profile detection in UPLC systems. It covers 190 nm and beyond with ±1 nm accuracy and ≤0.8 μL dispersion, distinguishing it by its analytical flow cell. It serves as the primary detector for acquiring full UV-Vis spectra in method development and routine analysis.
Full specifications
| Detector type | Photodiode Array (PDA) Detector |
| Wavelength range | 190 nm |
| Wavelength accuracy mfr | ±1 nm (via patented 1 Erbium filter) |
| Linearity range mfr | Deviation at 2.0 AU ≤5%, propylparaben, at 257 n |
| Optical resolution mfr | 1.2 nm |
| Digital resolution mfr | 0.6 nm/pixel to 500 nm |
| Drift mfr | ≤1.0 x 10 –3 /AU/hour/°C with analytical flow ce |
| Sampling rate mfr | Up to 80 points/s |
| Light source mfr | Prealigned, intelligent technology Deuterium lam |
| Dispersion volume mfr | ≤0.8 μL (Analytical flow cell); ≤1.5 μL (High se |
| Flow cell design mfr | Light-guiding UPLC intelligent technology flow c |
| Flow cell volume mfr | 500 nL (analytical flow cell) |
| Wetted materials mfr | 316 stainless-steel, fused silica, Teflon AF, PE |
| Pressure limit mfr | 1000 psi (all flow cells) |
| Path length mfr | 10 mm (analytical flow cell) |
| Width mfr | 34.3 cm (13.5 inches) |
| Height mfr | 19.4 cm (7.6 inches) |
| Depth mfr | 60.7 cm (23.9 inches) |
| Noise | ±3 μAU, 230 nm, 2 points/s, 2 s, wavelength compensation 310 to 410 nm, 90:10 water/methanol (2D channel with analytical flow cell) |
Description
Construction and Positioning
- Photodiode array (PDA) detector with a prealigned, intelligent deuterium lamp and a light-guiding UPLC intelligent flow cell. - Analytical flow cell: 500 nL volume, 10 mm path length, dispersion ≤0.8 μL; high-sensitivity flow cell (≤1.5 μL dispersion) is available but not part of this configuration. - Wetted materials: 316 stainless steel, fused silica, Teflon AF, PEEK. ### Performance Characteristics
- Wavelength range starts at 190 nm; wavelength accuracy ±1 nm using a patented erbium filter. - Noise: ±3 μAU at 230 nm (2 points/s, 2 s, wavelength compensation 310–410 nm, 90:10 water/methanol). - Linearity: deviation ≤5% at 2.0 AU (propylparaben, 257 nm). - Optical resolution 1.2 nm; digital resolution 0.6 nm/pixel up to 500 nm. - Drift ≤1.0 × 10⁻³ AU/hour/°C with analytical flow cell. - Sampling rate up to 80 points/s. ### Installation and Operation
- Pressure limit for all flow cells: 1000 psi. - Dimensions: 34.3 cm (W) × 19.4 cm (H) × 60.7 cm (D).
Integration & Operating Notes
Exceeding 1000 psi may damage the flow cell and compromise the light-guiding path. Ensure system pressure is regulated and monitored, especially when the detector is placed after high-pressure stages or when using narrow-bore tubing.
The analytical flow cell uses 316 stainless steel, fused silica, Teflon AF, and PEEK. Avoid solvents that attack these materials, such as strong acids or bases, to prevent leaks and baseline noise.
Selection & Integration Q&A
Choose this PDA for spectral profile detection, peak purity, and multi-wavelength acquisition. A single-wavelength detector is sufficient for routine quantitation at one wavelength and offers simpler method transfer.
The deuterium lamp is prealigned and intelligent; replacement is straightforward. No other maintenance schedule or consumables are specified in the provided facts. Refer to the manual for lamp replacement intervals and accessories.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions