Waters Alliance iS Bio HPLC System
Waters ·Made in USA · In Active Production

Waters Alliance iS Bio HPLC System

Waters Alliance iS Bio HPLC System — Chromatographs with Flow rate 10 milliliter/minute, Max pressure 82.74 megapascal.

The Alliance iS Bio HPLC System is a quaternary, inert flow-path HPLC system with a flow-through needle and MaxPeak HPS technology, supporting up to 10.0 mL/min and 12,000 psi. It accommodates one column up to 300 mm length and 8 mm I.D., with sample capacity for any three of six vial/plate formats. It is intended for bio-separation methods requiring pH 1–13 compatibility and low carryover (≤0.002% caffeine).

Flow rate
10 milliliter/minute
Max pressure
82.74 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 10.0 mL/min
Max pressure 12000 psi
Injection type mfr Flow-through needle
Sample flow path mfr Inert (MaxPeak HPS Technology)
Sample carryover mfr ≤0.002% [Caffeine] under UV conditions
Sample compartment mfr 4 to 40 °C
Pump options mfr Quaternary
Solvent conditioning mfr Passive pre-heating
Column tracking mfr Yes, with RFID enabled eConnect Waters Columns
Column capacity mfr 1 column up to 300 mm length, up to 8 mm interna
Maximum sample capacity mfr (any three of the following) 48-position, 2.00-m
Integrated Detector Options mfr TUV Detector, PDA Detector
External Detectors Supported mfr CAD, ACQUITY QDa II, ELS, RI, FLR, MALS
Informatics compatibility mfr Empower Chromatography Data System (CDS) (Featur
Dimensions, Weight & Mechanical Design
Height 742 mm
Weight 72.5 kg
Width 499 mm
Depth mfr 63.5 cm (24.99 in.)
Materials, Media & Process Compatibility
pH range mfr 1 to 13

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 12,000 psi system pressure

Operating above the 12,000 psi (≈82.7 MPa) limit risks seal, fitting, or column hardware failure and invalidates results. Verify backpressure at the maximum method flow (10.0 mL/min) with the installed column I.D. and length before routine use.

Factory Specification Sheet
Process Limits
Mobile-phase pH must stay within 1–13

The inert MaxPeak HPS flow path is rated for pH 1–13. Methods outside this window can degrade wetted materials and produce metal-ion interference or leaks; confirm buffer pH before priming the quaternary pump.

Factory Specification Sheet
Control & I/O
Empower CDS FR4 or later is required for control

The system is controlled through Empower Chromatography Data System Feature Release 4 on Windows 10 or later. Older CDS releases or unsupported operating systems will not provide instrument control or data acquisition; plan the informatics upgrade before installation.

Factory Specification Sheet

Selection & Integration Q&A

What external detectors and informatics does the system support for adjacent integration?

Integrated TUV or PDA detectors are available; external CAD, ACQUITY QDa II, ELS, RI, FLR, and MALS detectors are supported. Control and data acquisition require Empower CDS Feature Release 4 on Windows 10 or later.

What maintenance and consumables are required, and what is not included in base delivery?

Column tracking requires RFID-enabled eConnect Waters Columns; non-RFID columns need a manual log. No maintenance schedule, service intervals, cables, or consumables are specified in the provided facts; confirm the base-delivery scope with the vendor.

Dimensional Drawings & Technical Documentation

Technical documentation