Agilent 2100 Bioanalyzer System Maintenance and Troubleshooting Guide

Technical Reference Summary

Covered Models & Document Scope

Maintenance and troubleshooting for the Agilent 2100 Bioanalyzer system, based on 2100 Expert Software revision B.02.08 and newer. Covers instrument communication, 2100 Expert software, hardware diagnostics, DNA/RNA/Protein applications, electrode cartridge, chip priming station, instrument, vortexer, and spare parts. Diagnostic test applicability notes G2939A, G2938B and C instruments for the high voltage accuracy-on load test.

Section-by-Section Technical Breakdown

pp. 6-13
Essential Measurement Practices

Provides user-relevant handling hints for tools, chips, reagents, samples, the chip priming station, and the 2100 Bioanalyzer instrument. Emphasizes contamination prevention, proper pipetting, reagent equilibration, and chip handling within time limits.

  • Always wear gloves when handling chips; use non-filter pipette tips of adequate size.
  • Insert pipette tip to the bottom of the well when dispensing liquid to avoid bubbles.
  • Replace syringe with each new kit; perform chip priming station seal test monthly.
  • Keep reagents refrigerated at 4°C when not used for more than 1 hour; equilibrate to room temperature for 30 minutes before use.
  • Protect dye and gel-dye mixes from light.
  • Prepared chips must be used within 5 minutes.
  • For DNA and RNA assays, vortex chips for 1 minute using only the IKA vortexer.
  • Do not leave any wells empty; for DNA/RNA add 1 µL sample buffer to unused sample wells so total volume is at least 6 µL.
  • Clean electrodes daily with electrode cleaner and monthly with toothbrush and distilled water.
  • Clean focusing lens monthly or after liquid spill using isopropanol.
pp. 14-22
Troubleshooting the Instrument Communication

Describes how to verify PC-instrument communication, troubleshoot communication issues, change COM port settings, use a USB-to-serial adapter, and create a support package. Includes a flow chart and notes on license keys and status indicator behavior.

  • Status indicator lamp lights green if power is present and self-tests pass; green does not indicate PC communication.
  • Verify communication by opening/closing lid and observing icon change in 2100 Expert Software.
  • Ensure two licenses are registered: instrument control license and electrophoresis license.
  • Check COM port settings; laptop PCs have only one COM port.
  • If status indicator is red, power cycle instrument; if off and fan not running, replace fuses.
  • Agilent recommends direct serial connection; if unavailable, use Agilent USB-Serial Adapter cable (Part number 5188-8031, black cable) for software B.02.08 and newer.
  • Install USB-serial driver from SupportDriver/88-8031CDM21228.exe before connecting adapter.
  • Create support package via Help > Create support package; ZIP file saved to desktop.
pp. 23-25
Troubleshooting the 2100 Expert Software

Explains how to run the Installation Qualification Test to check for corrupted or missing software files. Provides steps to start the test, review results, and reinstall software if the test fails.

  • Start 2100 Expert Software and select Validation (Verification) context.
  • Select New Validation (Verification) > Installation Validation (Verification) > Software > [My PC Name] > Installation Qualification Test.
  • Select Installation Qualification Test check box and click Start.
  • If test fails, reinstall 2100 Expert Software using supplied UFD media.
  • If test continues to fail, contact Agilent Technologies.
pp. 26-30
Hardware Diagnostics

Lists available test chips and hardware diagnostic tests for electrophoresis mode, and describes the diagnostic test procedure. Tests should be run regularly or when incorrect measurements or error messages occur.

  • Test chips: Autofocus test chip (values printed on chip, reusable) and Electrode/Diode test chip (reusable).
  • Diagnostic tests include Electronic test, Fan test, Lid sensor test, Temperature test, Stepper motor test, Electrode/Diode test, High voltage stability test, High voltage accuracy test, High voltage accuracy-on load test (only for G2939A, G2938B and C instruments), Short circuit test, Optics test, Autofocus test, Laser stability test.
  • Short circuit test limits specify ambient temperature of 25°C and relative humidity ≤50%.
  • Diagnostic tests cannot be run while a chip run is in progress; instrument must be online.
  • Results stored as .xdy files in Agilent\2100 bioanalyzer\2100 expert\diagnosis; send to Agilent Technical Support if tests fail.
pp. 31-61
Troubleshooting the DNA Application

Provides symptom-based troubleshooting for DNA and High Sensitivity DNA assays, listing causes by probability and corresponding solutions. Covers issues such as quantitation errors, sizing errors, run aborts, chip detection, peak anomalies, baseline problems, and migration issues.

  • Symptoms include Residual Gel in Spin Filter after Centrifugation, Too High/Low Quantitation Results, Wrong Sizing Result, Run Aborted, Chip Not Detected, Additional Sample or Ladder Peaks, Spikes, Low Signal Intensity, Missing Peaks, Missing Upper Marker, Broad Peaks, Baseline Dips/Noise/Jumps, Wavy Baseline, Late Migration, Peak Tailing, Unexpected Run Time, Error Message: No data received since 5 seconds.
  • High Sensitivity DNA symptoms: Artefact Peaks, Split Peaks, Baseline Negative Dips.
  • Causes grouped into Most probable, Probable, Least probable.
  • Run Log in Log Book tab lists actions and errors during run.
  • Manual marker assignment possible via right-click on peak.
pp. 62-82
Troubleshooting the RNA Application

Provides symptom-based troubleshooting for RNA assays, listing causes and solutions. Section overview and symptoms are referenced in the table of contents.

  • Symptoms (RNA) section starts on page 64.
  • Refer to Kit Guides for assay-specific limits.
pp. 83-109
Troubleshooting the Protein Application

Provides symptom-based troubleshooting for Protein and High Sensitivity Protein assays, listing causes and solutions. Section overview and symptoms are referenced in the table of contents.

  • Symptoms (Protein) section starts on page 85.
  • Symptoms (High Sensitivity Protein) section starts on page 107.
pp. 110-125
Maintenance of the Electrode Cartridge

Describes maintenance procedures for the electrode cartridge for DNA, protein, RNA Nano, RNA Pico, and Small RNA assays, and how to clean the pin set. Includes cleaning instructions to prevent leak currents and contamination.

  • Clean electrodes daily using electrode cleaner.
  • Thoroughly clean electrodes monthly using a toothbrush and distilled water.
  • Clean electrodes with analysis-grade water and a toothbrush to address leak currents.
  • Do not force chip into instrument; electrode cartridge may be damaged when lid is closed.
pp. 126-135
Maintenance of the Chip Priming Station

Covers maintenance of the chip priming station including replacing the syringe, cleaning and replacing the syringe adapter, replacing the gasket, and performing a seal test to check performance.

  • Replace syringe with each new kit.
  • Check performance monthly by applying seal test.
  • Replace gasket and/or adapter if necessary.
  • Refer to Kit Guide for correct position of syringe clip and base plate.
pp. 136-143
Maintenance of the Agilent 2100 Bioanalyzer Instrument

Describes cleaning the focusing lens and changing the fuses. Lens cleaning should be done monthly or after any liquid spill using isopropanol.

  • Clean focusing lens once a month or after any liquid spill using isopropanol.
  • Replace fuses if status indicator is off and fan is not running; spare fuses come with instrument.
pp. 144-145
Maintenance of the Vortexer

Covers changing the vortexer adapter. Replacement adapters are available for MS-2 and MS-3 vortexers.

  • For MS-2 vortexers with 3 mounting screws, replacement part number is 5065-9966.
  • For MS-3 vortexers with 4 mounting screws, replacement adapters may be purchased from IKA (www.ika.de) with part number 3428300.
  • Press vortex adapter tightly on mount; adapter must not rock.
pp. 146-147
Spare Parts and Accessories

Provides an overview of spare parts and accessories for the 2100 Bioanalyzer system. Specific part numbers are referenced throughout the manual.

  • USB-Serial Adapter cable part number 5188-8031 (black cable) for software B.02.08 and newer.
  • Vortexer adapter part numbers: 5065-9966 (MS-2, 3 screws) and 3428300 (MS-3, 4 screws, from IKA).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Prepared chips must be used within 5 minutes. p.12 ↗ Delays between preparation and run cause reagent evaporation and poor results. whole series
For DNA and RNA assays, vortex chips for 1 minute using only the IKA vortexer. p.12 ↗ Inappropriate or insufficient vortexing leads to poor results; other vortexers are not acceptable. DNA and RNA assays
For DNA and RNA assays, add 1 µL of sample buffer to each unused sample well so total liquid volume in each well is at least 6 µL. p.12 ↗ Empty wells cause the assay to run improperly. DNA and RNA assays
Keep reagents and reagent mixes refrigerated at 4°C when not in use for more than 1 hour. p.10 ↗ Reagents left at room temperature may decompose, leading to poor measurement results. whole series
Allow all reagents and samples to equilibrate to room temperature for 30 minutes; mix and spin down prior to use. p.10 ↗ Cold or viscous reagents cause poor results; equilibration is required. whole series
For RNA assays, heat denature all RNA samples and RNA ladder for 2 minutes at 70°C before use. p.11 ↗ Required denaturation step for RNA assays. RNA assays
For protein assays, use 0.5 mL tubes for denaturation. p.11 ↗ Using larger tubes leads to poor results. Protein assays
Clean electrodes daily using the electrode cleaner and monthly using a toothbrush and distilled water. p.13 ↗ Prevents leak currents and contamination that cause baseline and signal problems. whole series
Clean the focusing lens once a month (or after any liquid spill) using isopropanol. p.13 ↗ Fingerprints or contamination on lens cause low signal intensity and baseline noise. whole series
Check the performance of the chip priming station by applying the seal test on a monthly basis. p.9 ↗ Ensures proper chip priming; replace gasket and/or adapter if necessary. whole series
Use only 350 µL of water in the cleaning chip. p.49 ↗ Excess water can cause leak currents due to wet cartridge. whole series
Ensure humidity in the room is below 70% at 15–27°C (59–81°F). p.49 ↗ High humidity can cause leak currents and wavy baseline. whole series
Short circuit test limits specify ambient temperature of 25°C and relative humidity ≤50%. p.28 ↗ Higher temperature or humidity could result in a leak current and test failure. whole series
High voltage accuracy-on load test is only for G2939A, G2938B and C instruments. p.28 ↗ This diagnostic test is not available on other instrument variants. G2939A, G2938B and C instruments
Use Agilent USB-Serial Adapter cable (Part number 5188-8031, black cable) for 2100 Expert Software version B.02.08 and newer. p.20 ↗ Required for USB connection; install driver from SupportDriver/88-8031CDM21228.exe before connecting. 2100 Expert Software B.02.08 and newer
For MS-2 vortexers with 3 mounting screws, replacement adapter part number is 5065-9966; for MS-3 vortexers with 4 mounting screws, use IKA part number 3428300. p.12 ↗ Correct adapter ensures proper vortexing and prevents poor results. MS-2 and MS-3 vortexers
Diagnostic test results are stored as .xdy files in Agilent\2100 bioanalyzer\2100 expert\diagnosis. p.30 ↗ Send .xdy files to Agilent Technical Support if tests fail. whole series
If status indicator is red, turn off power and turn on again; if off and fan not running, replace fuses. p.17 ↗ Basic power/communication troubleshooting step. whole series
Ensure two licenses are registered: instrument control license and electrophoresis license. p.17 ↗ Missing licenses prevent instrument communication. 2100 Expert Software
Do not touch the 2100 Bioanalyzer instrument during a run and never place it on a vibrating surface or near air-circulating instruments. p.13 ↗ Vibration and temperature changes cause baseline jumps, spikes, and sizing errors. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Run aborted on port x. Instrument error occurred on port x, Unusual high or low voltage or current was detected during the start phase of the on-chip analysis. p.37 ↗ Insufficient volume in well(s); air bubble at bottom of well; dirty electrodes; chip not properly primed; chip prepared with cold reagents; high voltage power supply defective. Check Kit Guide for liquid amounts; ensure all wells contain sufficient ladder, samples or buffer; insert pipette tip to bottom of well; remove large air bubbles; clean electrodes; prepare new chip and check priming station; allow reagents to warm to room temperature for 30 minutes; check high voltage power supply via hardware diagnostics. Logbook error message; marked wells indicate problematic wells.
Chip Not Detected p.38 ↗ No communication between instrument and PC; insufficient volume in well(s); chip not properly primed/air bubble; expired reagents; high voltage power supply defective. Check instrument communication; check Kit Guide for liquid amounts; prepare new chip and check priming station; prepare new chip with fresh reagents; check high voltage power supply via hardware diagnostics. Section 5, Symptoms (DNA)
Error Message: No data received since 5 seconds p.57 ↗ Not detailed in provided text. Not detailed in provided text. Listed in Symptoms (DNA) table of contents, page 57.
Residual Gel in Spin Filter after Centrifugation p.34 ↗ Gel filtered at insufficient g-value; cooled centrifuge used; gel too cool or viscous. Refer to Kit Guide for proper centrifuge settings; repeat centrifugation at room temperature; equilibrate reagents at room temperature for 30 minutes prior to use. Section 5, Symptoms (DNA)
Too High Quantitation Results p.34 ↗ Pipetting error during reagent mix preparation; chip pipetting error; insufficient vortexing; dye concentration too low; low or missing upper marker; loaded chip kept too long before run. Check dilution procedure and pipette calibration; prepare new chip and pipette to bottom of well; vortex chip for 1 minute with IKA vortexer; use correct dye concentration and warm dye to room temperature for 30 minutes protected from light; check Missing Upper Marker; use prepared chip within 5 minutes. Section 5, Symptoms (DNA)
Too Low Quantitation Results p.35 ↗ Pipetting error; chip pipetting error; insufficient vortexing; loaded chip kept too long; dye concentration too high; sample concentration too high. Check dilution procedure and pipette calibration; prepare new chip and pipette to bottom of well; vortex chip for 1 minute with IKA vortexer; use prepared chip within 5 minutes; use correct dye concentration and warm dye to room temperature for 30 minutes; use sample concentration according to Kit Guide. Section 5, Symptoms (DNA)
Wrong Sizing Result p.36 ↗ DNA ladder degraded; chip not properly primed/clogged priming station/wrong settings; chip contaminated; markers called incorrectly; loaded chip kept too long; no ladder in ladder well; vibration; ambient temperature changes >5°C; high voltage power supply defective. Check expiration date; prepare new chip and check priming station seal test; clean/replace syringe, gasket, adapter; check clip and base plate positions; wear powder-free gloves; do not touch chip underside or wells; clean electrodes; load chip immediately; manually assign lower marker; use chip within 5 minutes; prepare new chip with ladder; do not touch instrument during run; remove vibration devices; place in thermally stable environment; check high voltage power supply. Section 5, Symptoms (DNA)
Additional Sample or Ladder Peaks p.39 ↗ Chip or gel-dye mix contaminated with particles; sample degraded or contaminated; chip prepared with cold reagents; chip not properly primed; vibration; dye agglomerates; DNA ladder degraded. Prepare new chip with new gel-dye mix; wear powder-free gloves; do not touch chip underside or wells; clean electrodes; load chip immediately; always wear gloves; allow reagents to warm to room temperature for 30 minutes; check priming station; do not touch instrument during run; remove vibration devices; use correct dye concentration and warm/protect from light; take gel-dye mix from top of tube after centrifugation; check expiration date and use fresh DNA ladder. Section 5, Symptoms (DNA)
Spikes p.40 ↗ Vibration of instrument; chip or gel-dye mix contaminated with particles; dye agglomerates; chip prepared with cold reagents. Do not touch instrument during run; remove vibration devices; prepare new chip with new gel-dye mix; wear powder-free gloves; do not touch chip underside or wells; clean electrodes; load chip immediately; use correct dye concentration and warm/protect from light; take gel-dye mix from top of tube after centrifugation; allow reagents to warm to room temperature for 30 minutes. Section 5, Symptoms (DNA)
Low Signal Intensity p.41 ↗ Dye concentration too low; pipetting error; chip pipetting error; fingerprint on focusing lens or chip backside; insufficient vortexing; chip contaminated; vibration; autofocus failure. Use correct dye concentration and warm/protect from light; check dilution procedure and pipette calibration; pipette new chip to bottom of well; clean lens with isopropanol; do not touch chip underside; vortex chip for 1 minute with IKA vortexer; wear powder-free gloves; do not touch wells; clean electrodes; load chip immediately; do not touch instrument during run; remove vibration devices; check autofocus via hardware diagnostics. Section 5, Symptoms (DNA)
Missing Peaks p.43 ↗ Sample salt concentration too high; chip not properly primed; leak currents due to contaminated electrodes; laser broken; autofocus failure or high voltage power supply defective. Refer to Kit Guide for maximum salt limits; dilute samples with deionized DNase-free water; prepare new chip and check priming station; clean electrodes with analysis-grade water and toothbrush; perform Optics, Autofocus, and Laser Stability tests; check autofocus and high voltage power supply via hardware diagnostics. Section 5, Symptoms (DNA)
Missing Upper Marker p.44 ↗ Alignment of upper marker not set properly; upper marker digested by restriction enzymes; sample salt concentration too high. Manually assign upper marker; inactivate restriction enzymes by adding EDTA or heat per manufacturer's instructions; refer to Kit Guide for maximum salt limits and dilute samples with deionized DNase-free water if necessary. Section 5, Symptoms (DNA)
Broad Peaks p.45 ↗ Leak currents due to contaminated electrodes; chip not properly primed; dye concentration too high; genomic DNA or cDNA contamination. Clean electrodes with analysis-grade water and toothbrush; prepare new chip and check priming station; use correct dye concentration and warm to room temperature for 30 minutes; check DNA preparation procedure. Section 5, Symptoms (DNA)
Baseline Dips p.46 ↗ Sample concentration too high; sample impurities; sample salt concentration too high; dye concentration too low; autofocus failure. Use sample concentration according to Kit Guide; check DNA isolation protocol and clean up samples; refer to Kit Guide for maximum salt limits and dilute with deionized DNase-free water; use correct dye concentration and warm/protect from light; check autofocus via hardware diagnostics. Section 5, Symptoms (DNA)
Baseline Noise p.47 ↗ Chip contaminated; fingerprint on focusing lens or chip backside; autofocus failure or high voltage power supply defective. Wear powder-free gloves; do not touch chip underside or wells; clean electrodes; load chip immediately; clean lens with isopropanol; check autofocus and high voltage power supply via hardware diagnostics. Section 5, Symptoms (DNA)
Baseline Jumps p.48 ↗ Vibration of instrument; instrument lid touched during run; laser defective. Remove vibration devices from bench; do not touch instrument during run; check laser via hardware diagnostics. Section 5, Symptoms (DNA)
Wavy Baseline p.49 ↗ Leak currents due to contaminated electrodes; leak currents due to liquid spillage on top of chip; dye concentration too low; chip not properly primed; leak currents due to wet cartridge; ambient temperature changes >5°C; high voltage power supply defective. Clean electrodes; prepare new chip and lower vortexing speed or mix manually; use correct dye concentration and warm to room temperature for 30 minutes; check priming station; use only 350 µL water in cleaning chip; ensure humidity below 70% at 15–27°C; place in thermally stable environment; check high voltage power supply. Section 5, Symptoms (DNA)
Late Migration p.50 ↗ Chip not properly primed; leak currents due to contaminated electrodes; leak currents due to liquid spillage; loss of gel separation properties; dye concentration too high; vortex speed too high; genomic DNA or high molecular weight DNA contamination; vortex adapter not connected tightly; ambient temperature changes >5°C. Prepare new chip and check priming station; clean electrodes; prepare new chip and lower vortexing speed or mix manually; check Kit Guide for proper storage and use gel-dye mix within indicated time; use correct dye concentration; vortex chip for 1 minute with IKA vortexer at setpoint; check DNA isolation protocol; press vortex adapter tightly on mount; replace adapter if needed; place in thermally stable environment. Section 5, Symptoms (DNA)
Peak Tailing p.52 ↗ Sample salt concentration too high. Refer to Kit Guide for maximum salt limits; dilute samples with deionized DNase-free water if necessary. Section 5, Symptoms (DNA)
Unexpected Run Time p.53 ↗ Lower and/or upper markers are called incorrectly. Turn off alignment and check which bands are correct lower and upper markers; manually assign markers. Section 5, Symptoms (DNA)