Product overview, features and intended use of the spectrometer
Procedurepp. 9-22
Components, battery installation, software startup and charging procedures
Qualitypp. 16-20
Diagnostic values and performance validation checks
Procedurepp. 23-34
Steps for collecting spectra and using ATR internal reflectance
Procedurepp. 35-36
Routine maintenance and care instructions
Orderingpp. 37-38
List of replaceable parts and ordering information
Specpp. 39-41
Technical specifications and environmental operating conditions
p. 19table, 3 rows
p. 20table, 1 rows
p. 18table, 6 rows
p. 37table, 15 rows
p. 13table, 8 rows
Table 3. Minimum performance for the Signal-to-Noise test
Sample interface
Specification @ 2500 cm-1
Specification @ 1000 cm-1
ATR
2500
6000
External reflectance
1500
2800
Grazing angle
1500
2800
Table 4. Minimum performance for the Stability test
Time
Specification @ 3000 cm-1
Specification @ 1000 cm-1
30 min
2%
1%
Table 2. Diagnostic values
Value
Optimal (green)
Marginal (yellow)
Critical (red)
Comment
Energy (centerburst)
29,000–21,000
>29,000 or <21,000
> 31,000 or < 18,000
Indicates the overall alignment and proper gain adjustment for the system.
Battery life
AC or >30 min
30–5 min
<5 min
Source
1.9 A
>2.2 or <1.6
>2.5 or <1.0
Indicates a problem with the source control voltage or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal
12,000–4,000
>12,000 or <4,000
>15,000 or < 3,000
Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature
35–44
<35 or >44
<30 or >48
Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature
10–75
>75
>80
Indicates the ambient temperature over the specified range.
5. Spare Parts
Part number
Description
0024 - 001
Agilent 4200 FlexScan FTIR system with external reflectance
0024 - 002
Agilent 4200 FlexScan FTIR system with diffuse reflectance
0024 - 003
Agilent 4200 FlexScan FTIR system with grazing angle
0024 - 004
Agilent 4200 FlexScan FTIR system with diamond ATR
0023 - 300 -
Diffuse gold reflectance reference cap for diffuse reflectance
0023-301
Polystyrene frequency reference cap for diffuse reflectance
0023 - 304
Diffuse gold reflectance reference cap for external reflectance
0023 - 305
Polystyrene frequency reference cap for external reflectance
0023 - 307
Mirrored stainless steel reflectance reference cap for external reflectance
0023 - 520
ATR sample press
800 - 2002
Handheld PC (PDA)
800 - 2003
Handheld PC docking station (includes USB and power cables)
800 - 2004
Replacement handheld PC battery
800 - 2005
Replacement lithium-ion battery
430 - 0018
USB cable
For sampling accessories, upgrades, service contracts, repair and refurbishment services, contact Agilent.
Table 1. Power switch LED status
Status
LED color
Action (% duty cycle)
System OFF
No LED illumination
N/A
System OFF
Red
100%
System startup
Red / green
Green 0.5 s on / red 0.5 s on
System ON
Green
100%
Low battery
Green
50% (0.5 s on / 0.5 s off)
Critically low battery
Red
50% (0.5 s on / 0.5 s off)
Dead battery
No LED illumination
N/A
Firmware update
Red / green
Green blinks rapidly twice then red blinks rapidly twice
Engineering Facts
14 facts · every claim linked to a page
Safety
The instrument is not intrinsically safe and must only be used in atmospheres tested for flammable materials.
Determines whether the analyzer can be deployed in hazardous or potentially explosive environments.
Source
“4200 FlexScan FTIR systems are NOT intrinsically safe. Use the system only in atmospheres that have been tested for flammable materials.”
Selection
The 45-degree specular probe is intended for films thicker than 1 micron, while the grazing angle probe is for films thinner than 1 micron.
Guides probe selection based on expected coating or film thickness.
Source
“The 45 degree probe is optimal for analysis of thin films greater than 1 micron in thickness.”
Selection
The grazing angle probe should be chosen when nanometer-level contamination must be detected.
Defines the probe choice for ultra-thin contamination analysis.
Source
“The grazing angle probe should be used when very low level contamination needs to be analyzed (nanometer thickness levels).”
Installation
After delivery, the shipping case must be allowed to reach room temperature for several hours before opening to avoid condensation.
Prevents condensation damage during initial setup and installation.
Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation
The instrument must be kept away from hot surfaces and electromagnetic interference sources.
Affects placement and mounting location to avoid measurement errors and damage.
Source
“The instrument must be kept away from hot surfaces and any sources of electromagnetic interference.”
Operation
The battery door must remain closed and locked while the system is in use.
Ensures safe and reliable battery operation during field measurements.
Source
“The door must remain closed and locked while the system is in use.”
Operation
The power button must be held for two seconds to turn the system on or off.
Prevents accidental power-off and clarifies startup procedure.
Source
“In order to turn the system on or off, hold the button down for two (2) seconds.”
Operation
The instrument must be connected via Bluetooth for MicroLab Mobile to launch.
Critical for establishing communication between the handheld PC and spectrometer.
Source
“The instrument must be connected via Bluetooth for the MicroLab Mobile software to launch.”
Performance
The system should be warmed up for at least 30 minutes before running performance validation tests.
Ensures valid performance validation results and stable measurements.
Source
“The 4200 FlexScan system should be warmed up for at least 30 minutes before conducting any of these tests.”
Operation
For reflectance measurements, the sample absorbance should be between 0.01 and 2 AU.
Helps verify proper sample contact and measurement quality in real time.
Source
“For 4200 FlexScan FTIR applications using reflectance, the value should be between 0.01 and 2 AU.”
Selection
The diamond ATR accepts samples with a pH range from 1 to 14.
Defines chemical compatibility limits for the ATR sampling interface.
Source
“Diamond can accept samples with a pH range from 1 to 14, which means hard or abrasive samples and even strong acids can be safely analyzed.”
Operation
The ATR diamond should not contact the sample or any substance other than air during background measurement.
Ensures an accurate background spectrum for ATR analysis.
Source
“The diamond should not be in contact with the sample or any substance other than the air in the atmosphere during the background measurement.”
Operation
The solid sample press delivers 15 pounds of force and is not for use with liquids, gels, or pastes.
Prevents misuse of the press and ensures proper contact for solid and powder samples.
Source
“The press is designed to deliver 15 pounds of force to the sample or over 2500 pounds per square inch of pressure. It is not for use with liquids, gels or pastes”
Maintenance
The outside of the unit may be cleaned with a damp, soft cloth and mild detergent, but must not be immersed in water.
Defines safe cleaning procedure and warns against water ingress into the electrical compartment.
Source
“Use a damp, soft cloth and a mild detergent to wipe down the outside of the unit. Do not completely immerse the 4200 FlexScan in water.”