571 Ionization Gauge User Manual

Contents

  1. Warning p. 2
    Legal and safety notices for the manual
  2. p. 3
    Table of contents of the manual
  3. Overview pp. 4-5
    General instructions for using the gauge
  4. Overview p. 6
    Purpose and structure of the manual
  5. Overview p. 7
    Terms and definitions used in the document
  6. Warning p. 8
    Explanation of warning symbols used
  7. Warning p. 10
    Safety precautions and hazards
  8. Warning p. 11
    Use with combustible gases and mixtures
  9. p. 12
    Disposal instructions for the equipment
  10. Services p. 13
    Service and support information
  11. Spec p. 14
    Technical data and specifications
  12. Overview p. 15
    Description and principle of the gauge
  13. Config p. 16
    Electrical connection details and pinout
  14. Procedure p. 17
    Operating instructions for the gauge
  15. Spec p. 18
    Detailed technical specifications
  16. Procedure p. 20
    Mounting and installation instructions
  17. Procedure p. 21
    Instructions for using the gauge tube
  18. Spec p. 22
    X-ray limit and measurement range notes
  19. Spec p. 23
    Correction factors for different gases
  20. Ordering p. 25
    Ordering information for accessories and spares
  21. p. 31
    Customer care contact information
  22. p. 18 table, 9 rows
  23. p. 14 table, 11 rows
  24. p. 17 table, 3 rows
  25. p. 25 table, 4 rows
  26. p. 19 table, 5 rows
  27. p. 20 table, 5 rows

Table 3 Physical Specifications

CharacteristicSpecification
EnvelopeNonex (all models)
GridTungsten “Non-Sag”, 0.025 inch diameter
FilamentHairpin yttria-coated Iridium Filament
CollectorTungsten, 0.010 inch diameter
Shield CoatingPlatinum, internally connected to filament
Base LeadsSoft nickel, 0.060 inch diameter
Collector LeadSoft nickel, 0.040 inch diameter
Shipping Weight3 lbs. (6.6 kg)
Internal Volume220 cc (not including tubulation)

Table

РазделСтраница
Description of the 571 Ionization Gauge15
Electrical Connections16
Operation17
Sensitivity17
Technical Specifications18
Mounting20
Use of The Gauge Tube21
Degassing21
X- ray Limit22
Gas Correction Factors23
Accessories and spare parts25

Оглавление раздела Technical Information; заголовок таблицы в тексте не напечатан.

Table 1 Gas Sensitivity Parameters

GasGauge Constant S S = i+/i- x 1/PEmission Setting for 0.01 Amp/Torr sensitivity (10 uA/micron)
Helium1.5 per Torr6.7 mA
Nitrogen10 per Torr1.0 mA
Air10 per Torr1.0 mA

where: S = Sensitivity • Torr; i+ = Ion current, Amps; i- = Grid (electron) current, Amps; P = Partial pressure of the gas, Torr

Table 8 Accessories and spare parts

Part NumberDescription
L7430301Glass ion gauge connector kit
R32443010Cable, glass BA gauge standard, non-bakeable, XGS-600 gauge controller, 10 ft
R32443025Cable, glass BA gauge standard, non-bakeable, XGS-600 gauge controller, 25 ft
R32443050Cable, glass BA gauge standard, non-bakeable, XGS-600 gauge controller, 50 ft

Table 5 Degassing Ratings

Resistance Heating*Electron Bombardment
Grid Voltage6.3 VAC
Grid to Filament Voltage+700 VDC
Grid Current8.7 Amps100 mA DC
Grid Temperature1200 °C1200 °C max
Filament Temperature1400 °C

* All values in the Resistance Heating column are absolute maximum ratings. If these values are exceeded, sagging of the grid can occur. Electron Bombardment: 50 W absolute maximum

Table 6 NW25/NW40 KF and 2.75 CF flange Option

FlangeAB
2.75 in DN 40 ConFlat flange69.3 (2.73)25.4 (1.00)
NW25 KF40.1 (1.58)25.4 (1.00)
NW40 KF55.1 (2.17)25.4 (1.00)
1.00 in tube-25.4 (1.00)
0.75 in tube-19.3 (0.75)

Engineering Facts

14 facts · every claim linked to a page
Selection

The gauge covers a wide pressure range from 1 mTorr down to 4x10-10 Torr, making it suitable for general-purpose vacuum measurement.

Defines the application envelope for selecting this gauge versus alternatives.

Source
“The 571 is useful as a general purpose vacuum gauge over the pressure range of 1 mTorr down to 4x10-10 Torr.”
Safety

The controller and vacuum system must be separately grounded to a common ground; a ground wire between chamber and controller chassis is unsafe.

Improper grounding can expose personnel to lethal voltages up to 1000 V.

Source
“Check that your Ion Gauge Controller and vacuum system are separately grounded to a common ground. Placement of a ground wire between the vacuum chamber and the controller chassis is not safe; large continuous currents could flow through it.”
Installation

The system ground must be verified to support at least 10 A before proceeding.

Ensures the grounding path can carry fault currents safely.

Source
“Before proceeding, test the system ground to be sure that it is complete and capable of supporting at least 10A.”
Safety

The gauge is not intrinsically safe and must not be turned on when combustible gases or mixtures are present.

Hot filament and high voltage can ignite combustible gas mixtures, causing injury and equipment damage.

Source
“As with all ionization gauges, this device is not intrinsically safe. Exercise extreme care when using this vacuum gauge while pumping or backfilling a system or in any other system condition which contains combustible gases or mixtures.”
Installation

Mount the gauge as close as possible to the measurement location through a large tubulation, preferably 1 inch OD, to avoid gauge pumping effects.

Low conductance causes the gauge to read lower than system pressure, especially below 10-8 Torr.

Source
“In general, the gauge tube should be mounted as close as possible to the desired location through a large tubulation (1" OD recommended).”
Operation

The tube can operate continuously in argon at 1x10-1 Torr without affecting filament life, but tube life in air may be shortened above 1x10-3 Torr.

Informs allowable operating pressure and gas-dependent lifetime limits.

Source
“When used in argon, the tube can operate continuously at 1x10-1 Torr with no effect on filament life. Tube life in air may be shortened above 1x10-3 Torr”
Maintenance

Do not bake the gauge at pressures above 10-5 Torr to avoid oxidation of gauge elements.

Prevents permanent degradation that would prevent reaching ultrahigh vacuum.

Source
“High temperature baking of the gauge should not be attempted at pressures above 10-5 to avoid oxidation of the gauge elements which can make it very difficult to attain ultrahigh vacuum pressure readings.”
Maintenance

Electron bombardment degassing applies up to 700 VDC to the grid connections, which can be lethal.

Requires appropriate precautions during degassing maintenance.

Source
“If electron bombardment degassing is employed, high voltages (up to 700 VDC maximum) will be applied to the grid connections; such high voltages can kill.”
Performance

The X-ray limit is approximately 2x10-10 Torr for nitrogen, and readings near this value should not be considered unequivocal.

Defines the lower useful measurement limit and uncertainty near it.

Source
“Therefore, as the typical value of X-ray limit is approached (2x10-10 Torr nitrogen equivalent), readings should not be considered unequivocal.”
Operation

The gauge is composition-sensitive; readings are expressed as equivalent pressure of one gas, usually nitrogen.

Requires gas correction factors when measuring gases other than nitrogen.

Source
“All ionization gauges are composition-sensitive and pressure readings derived from these gauges can only be expressed in terms of equivalent pressure of one gas, usually nitrogen.”
Maintenance

Do not use silicone oil or silicone grease on vacuum equipment; use powder-free butyl or polycarbonate gloves.

Prevents contamination that would degrade vacuum performance.

Source
“Do not use silicone oil or silicone grease. Use powder-free butyl or polycarbonate gloves to prevent skin oils from getting on vacuum surfaces.”
Maintenance

Do not clean aluminum parts with Alconox because it is incompatible with aluminum and causes damage.

Avoids corrosive damage to aluminum vacuum components.

Source
“Do not clean any aluminum parts with Alconox®. Alconox is not compatible with aluminum and will cause damage.”
Selection

The yttria-coated iridium filament operates at lower temperature than tungsten, resulting in less outgassing and resistance to accidental overpressure.

Reduces contamination and improves robustness for vacuum system selection.

Source
“Its yttria-coated Iridium Filament operates at a lower temperature than tungsten resulting in less outgassing. In addition, it will liberate gas during occasional accidental overpressure without immediate failure.”
Operation

Use a proper gauge controller that does not limit the broad range of the gauge or exceed maximum ratings.

Ensures full measurement range and prevents damage from incompatible controllers.

Source
“It is important to use a proper gauge controller that does not limit the broad range of the gauge, or exceed any of the maximum ratings noted in this manual.”