Agilent Jet Stream Thermal Gradient Focusing Technology

Technical Reference Summary

Covered Models & Document Scope

Technical Note describing the Agilent Jet Stream thermal gradient focusing technology for electrospray LC/MS ion sources; covers recommended operating parameters and sensitivity gains across flow rates 50 µL/min to 2.5 mL/min (optimal 0.25–2.0 mL/min), demonstrated on an Agilent 6530 Q-TOF LC/MS system. No model/size/ordering variants are defined.

Technical Data

Parameter Value Condition
Sheath gas flow 11 mL/min p.4 ↗ optimal, flow rates 0.25–2.0 mL/min
Sheath gas temperature 350 °C p.4 ↗ optimal, flow rates 0.25–2.0 mL/min
Nozzle voltage 600 V p.4 ↗ optimal, flow rates 0.25–2.0 mL/min
Nebulizer pressure 30 psi p.4 ↗ optimal, flow rates 0.25–2.0 mL/min
Chamber voltage 4 kV p.4 ↗ optimal, flow rates 0.25–2.0 mL/min
Sensitivity gain 5-10x x p.4 ↗ flow rates 50 µL/min to 2.5 mL/min
Spray start temperature 25 °C p.2 ↗ time-lapse imaging
Spray stop temperature 350 °C p.2 ↗ time-lapse imaging

Operating Limits

  • Flow rate range with sensitivity gain: 50 µL/min to 2.5 mL/min p.4 ↗
  • Flow rate range for greatest gains: 0.25-1.0 mL/min p.4 ↗
  • Flow rate range for optimal parameters: 0.25 to 2.0 mL/min p.4 ↗