Agilent 7820A GC System

Contents

  1. Overview pp. 1-2
    Introduces the 7820A GC system and its positioning
  2. Overview pp. 4-5
    Describes ease of use, EPR and manual operation features
  3. Config p. 6
    Lists software choices available for the system
  4. Spec p. 7
    Highlights uncompromising analytical performance characteristics
  5. p. 8
    Provides contact and further information resources
  6. p. 4 table, 4 rows

Retention time and area repeatability split/splitless inlet and FID

CompoundRetention Time (min) 24 ˚CRetention Time (min) 30 ˚CPeak area (pA) 24 ˚CPeak area (pA) 30 ˚C
C133.78433.782045.214344.5105
C144.42134.419045.579044.8429
C155.03695.0344457.8670449.3625
C165.62385.6211456.6820449.5554

Engineering Facts

14 facts · every claim linked to a page
Selection

The system supports split/splitless, purged packed, and programmable cool on-column inlets, so inlet choice must match column type and diameter.

Determines whether the chosen column and application can be run on this GC without extra hardware.

Source
“Split/splitless (SSL) for large bore and all capillary columns”
Selection

The PCOC inlet is only specified for columns of 0.250 mm id or larger, which constrains column selection.

Prevents selecting narrow-bore columns that the on-column inlet cannot accommodate.

Source
“Programmable cool on-column (PCOC) for columns >= 0.250 mm id”
Selection

The micro-ECD is not offered in Japan, so regional availability must be checked before specifying that detector.

Affects detector configuration and ordering for Japanese installations.

Source
“Not available in Japan.”
Selection

The GC can automatically control up to two heated valves, enabling valve-based chemical and hydrocarbon processing setups.

Defines the maximum valve automation capacity when planning a valve-based analysis.

Source
“the 7820A GC system can automatically control up to two heated valves”
Operation

EPR allows manual electronic adjustment of pressures and flows but cannot store setpoints in a method, unlike EPC.

Critical for deciding between EPR and EPC when method reproducibility and automation are required.

Source
“There are no pressure or flow set points that can be entered or saved to a method.”
Operation

EPR includes ambient temperature and pressure compensation, improving retention time and baseline stability versus manual pneumatics.

Justifies EPR over manual pneumatics where drift from ambient changes is a concern.

Source
“EPR has built-in ambient temperature and pressure compensation”
Installation

A guided system setup lets users get the instrument running without an Agilent service representative visit.

Reduces installation cost and planning for on-site service support.

Source
“a guided system set-up that helps get you up and running without a visit by an Agilent service representative”
Operation

The 5-button keypad and absence of gauges or manual gas knobs reduce operator errors, suiting inexperienced users.

Supports the choice for labs with infrequent or untrained operators.

Source
“Because there are no gauges or manual gas knobs, errors are minimized.”
Maintenance

Built-in self-diagnostics and real-world design features make the system easy to maintain.

Informs expected maintenance effort and service planning.

Source
“with convenient, real-world design features and built-in self-diagnostics, the 7820A GC is also easy to maintain”
Performance

Full EPC on all inlets and detectors improves retention time and peak area repeatability by keeping setpoints constant.

Key performance basis for choosing EPC over manual or EPR pneumatics.

Source
“Full electronic pneumatics control (EPC)—available on all inlets and detectors—ensures a better repeatability of retention times and peak areas.”
Performance

The FID's wide dynamic range allows quantification of both very high and very low concentrations in one run, reducing sample prep.

Guides detector selection for samples with wide concentration spreads.

Source
“Wide dynamic response range enhances accuracy and minimizes sample prep requirements”
Selection

The 7650A autoinjector holds up to fifty 2 mL vials, defining sample capacity for unattended operation.

Determines throughput and walk-away time when configuring automation.

Source
“The Agilent 7650A autoinjector provides capacity for up to fifty 2 mL vials”
Compliance

OpenLAB CDS editions allow optional integration with OpenLAB ECM for central data storage, GLP, and compliance features.

Important for regulated labs needing data integrity and compliance support.

Source
“allow optional integration with OpenLAB ECM for central data storage, GLP, and compliance features”
Operation

Using nitrogen instead of helium as carrier gas dramatically reduces cost per sample in ECD pesticide analysis.

Influences carrier gas supply planning and operating cost.

Source
“using nitrogen in place of helium as the carrier gas dramatically reduces cost per sample”