Rapid and Accurate Forensics Analysis using High Resolution All
Technical Reference Summary
Covered Models & Document Scope
Application note describing the Agilent All Ions MS/MS approach for high resolution accurate mass screening of forensic samples using TOF or Q-TOF LC/MS instruments; covers HPLC conditions, ion source parameters, MS parameters (MS-mode and All Ions MS/MS mode), and sample preparation for a 30-compound test standard and spiked blood sample.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| UHPLC column | Acquity UPLC, HSS T3, 2.1 mm × 150 mm, 1.8 µm p.2 ↗ | |
| Mobile phase A | 10 mM ammonium formate + 0.05% formic acid in water p.2 ↗ | |
| Mobile phase B | 0.05% formic acid in acetonitrile p.2 ↗ | |
| Flow rate | 0.5 mL/min p.2 ↗ | |
| Stop time | 12.0 min p.2 ↗ | |
| Gas temperature | 300 °C p.2 ↗ | |
| Gas flow | 6 L/min p.2 ↗ | |
| Nebulizer | 35 psi p.2 ↗ | |
| Sheath gas temperature | 375 °C p.2 ↗ | |
| Sheath gas flow | 10 L/min p.2 ↗ | |
| Scan range (MS-mode) | 50–1,000 amu p.2 ↗ | MS-mode |
| Scan rate (MS-mode) | 1.5 spectra/s p.2 ↗ | MS-mode |
| Scan segments (All Ions MS/MS mode) | Three scan segments at 0 eV, 20 eV, and 40 eV p.2 ↗ | All Ions MS/MS mode |
| Scan range (All Ions MS/MS mode) | 50–1,000 amu p.2 ↗ | All Ions MS/MS mode |
| Scan rate (All Ions MS/MS mode) | 3 spectra/s p.2 ↗ | All Ions MS/MS mode |
| Internal reference | purine (m/z 121.0509) and HP 921 (m/z 922.0098) p.2 ↗ | Automatic internal reference mass correction |
| Injection volume | 2 µL p.2 ↗ | |
| Sample preparation | liquid/liquid extraction with acetonitrile/ethanol 90/10 (250 mg blood + 500 µL extraction solvent) p.2 ↗ |