Sensitive Determination of Polycyclic Aromatic Hydrocarbons in Tap Water
Technical Reference Summary
Covered Models & Document Scope
Application Note describing an online SPE UHPLC method for determination of 16 PAHs in drinking water on an Agilent 1200 Infinity Series Online SPE system (Agilent 1260 Infinity LC with autosampler); no product model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sample volume required | less than 2 mL p.2 ↗ | |
| Overall run time (enrichment + analytical run) | 30 minutes p.2 ↗ | |
| Number of PAHs monitored | 16 p.2 ↗ | |
| Autosampler sample capacity | 2 × 18, 6-mL vials p.2 ↗ | Agilent 1260 Infinity autosampler |
| Fluorescence detector acquisition | Multisignal acquisition, set at λex= 260 nm p.3 ↗ | Agilent 1290 Infinity fluorescence detector |
| Retention time RSD | 0.011–0.090 % p.4 ↗ | n = 7, 16 PAHs |
| Area RSD | 0.3–6.0 % p.4 ↗ | n = 7, 16 PAHs |
| Linearity R² | 0.99834–0.99996 p.5 ↗ | 16 PAHs |
| Limit of detection (LOD, S/N = 3) | 0.0003–0.0233 µg/L p.5 ↗ | 16 PAHs |
| Limit of quantification (LOQ, S/N = 10) | 0.0009–0.0775 µg/L p.5 ↗ | 16 PAHs |
| LOD range | 0.2–23 ng/L p.5 ↗ |
Operating Limits
- Minimum sample volume sufficient for analysis: less than 2 mL p.2 ↗
- Overall run time: 30 minutes p.2 ↗