Analysis of Arsine and Phosphine in Ethylene and Propylene Using the
Technical Reference Summary
Covered Models & Document Scope
Agilent Arsine Phosphine GC/MS Analyzer, model G3440B with option SP1 7890-0667, comprising Agilent 7890B GC and Agilent 5977B Series GC/MSD System HES with hydrogen cleaning; application for arsine and phosphine (plus H2S, COS) in ethylene and propylene.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analyzer model | G3440B p.2 ↗ | with option SP1 7890-0667 |
| Gas chromatograph | 7890B GC p.2 ↗ | |
| Mass spectrometer detector | 5977B Series GC/MSD System HES p.2 ↗ | with hydrogen cleaning |
| Column | 120 m × 0.32 mm, 8.0 µm Select Olefins p.2 ↗ | |
| Carrier gas | Helium, constant flow at 1.2 mL/min p.2 ↗ | |
| Oven program | 35 °C p.2 ↗ | for 22 minutes |
| Hydrogen flow into source | 150 µL/min p.2 ↗ | H2 at 40 psi |
| Linearity (R²), Phosphine (PH3) | 0.9996 p.3 ↗ | |
| Linearity (R²), Hydrogen sulfide (H2S) | 0.9995 p.3 ↗ | |
| Linearity (R²), Arsine (AsH3) | 0.9999 p.3 ↗ | |
| Linearity (R²), Carbonyl sulfide (COS) | 0.9987 p.3 ↗ | |
| RSD, Phosphine (PH3) | 5.14 % p.3 ↗ | |
| RSD, Hydrogen sulfide (H2S) | 3.96 % p.3 ↗ | |
| RSD, Arsine (AsH3) | 0.62 % p.3 ↗ | |
| RSD, Carbonyl sulfide (COS) | 5.06 % p.3 ↗ | |
| IDL, Phosphine (PH3) | 0.715 ppb p.3 ↗ | |
| IDL, Hydrogen sulfide (H2S) | 0.456 ppb p.3 ↗ | |
| IDL, Arsine (AsH3) | 0.063 ppb p.3 ↗ | |
| IDL, Carbonyl sulfide (COS) | 0.575 ppb p.3 ↗ |
Operating Limits
- Detection level (IDL) for arsine: 0.063 ppb p.3 ↗
- Detection level (IDL) for phosphine: 0.715 ppb p.3 ↗
- Detection level (IDL) for hydrogen sulfide: 0.456 ppb p.3 ↗
- Detection level (IDL) for carbonyl sulfide: 0.575 ppb p.3 ↗