Maintaining Sensitivity and Reproducibility with the
Technical Reference Summary
Covered Models & Document Scope
Application Note: analysis of approximately 200 pesticides in organic honey extract on an Agilent 7010A Triple Quadrupole GC/MS with Agilent 7890B GC and 7693B Autosampler, comparing operation with and without the Agilent JetClean self-cleaning ion source (continuous H2 flow 0.13 mL/min).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Continuous H2 flow into MS source (JetClean) | 0.13 mL/min p.1 ↗ | JetClean in use |
| Oven temperature | 310 °C p.2 ↗ | MS acquisition method |
| Spike level for MDL determination | 2.5 ppb p.1 ↗ | honey extract, 10 replicates |
| Confidence level for MDL calculation | 99 % p.1 ↗ | |
| Average MDL without JetClean | 0.170 ppb p.1 ↗ | without JetClean |
| Average MDL with JetClean | 0.147 ppb p.1 ↗ | with JetClean |
| Inlet liner | 4 mm ultra inert, splitless, single taper, glass wool liner (p/n 5190-2293) p.2 ↗ | Multimode Inlet (MMI) |
| Analytical column | Agilent J&W DB-5ms Ultra Inert, 15 m × 0.25 mm, 0.25 µm (p/n 19091S-431 UI) p.2 ↗ | two columns coupled via purged ultimate union |