Induztr
Engineering Digest Document Viewer Related Models

In-depth Peptide Mapping with Iterative MS/MS Acquisition on the Agilent 6545XT AdvanceBio LC/Q-TOF

Technical Reference Summary

Covered Models & Document Scope

Application note describing an LC/MS/MS peptide mapping workflow for CHO-cultured human IgG1 mAb using the Agilent AssayMAP Bravo, Agilent 1290 Infinity II LC, Agilent 6545XT AdvanceBio LC/Q-TOF with Iterative MS/MS, and MassHunter BioConfirm B.09.00; covers 15-minute and 30-minute gradients and 0.6 µg or 2 µg sample loading.

Technical Data

Parameter Value Condition
Analytical column AdvanceBio Peptide Mapping, 2.1 × 150 mm, 2.7 µm (p/n 653750-902) p.2 ↗
Column temperature 60 °C p.2 ↗
Autosampler temperature 4 °C p.2 ↗
Solvent A 0.1 % formic acid in water p.2 ↗
Solvent B 0.1 % formic acid in 90 % acetonitrile p.2 ↗
Flow rate 0.4 mL/min p.2 ↗
Gas temperature 325 °C p.2 ↗ 6545XT AdvanceBio LC/Q-TOF
Sample loading amount 0.6 µg p.2 ↗ per analysis
MS1 mass tolerance 10 ppm p.2 ↗ BioConfirm search
MS2 mass tolerance 30 ppm p.2 ↗ BioConfirm search
Peptide length limit 5 to 60 amino acids p.2 ↗ BioConfirm search
False discovery rate 0.1 % p.2 ↗ peptide-spectrum match filter

Operating Limits

  • Column temperature: 60 °C p.2 ↗
  • Autosampler temperature: 4 °C p.2 ↗
  • Maximum missed cleavages: 2 p.2 ↗
  • Peptide length range: 5 to 60 amino acids p.2 ↗
  • False discovery rate: 0.1 % p.2 ↗