Quantification of Crystallinity Using Transmission Raman Spectroscopy
Technical Reference Summary
Covered Models & Document Scope
Application note covering the Agilent TRS100 Raman system for bulk quantification of crystalline API content (0–9.4 % w/w) in spray dried solid dispersions; comparison with pXRD and ssNMR.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Limit of detection (LOD) | 0.9 % w/w p.2 ↗ | TRS100, crystalline API in amorphous solid dispersion |
| Crystalline content range analyzed | 0–9.4 % w/w p.2 ↗ | nine powder mixtures |
| Correlation fit (R2) | 0.99 p.2 ↗ | PLS chemometric model |
| Data acquisition time | 1–5 minutes p.2 ↗ | TRS |
| Sensitivity (LOD) | 0.2–1 % p.2 ↗ | TRS |
| Sensitivity (LOD) | 2–5 % p.2 ↗ | pXRD |
| Sensitivity (LOD) | 0.3–1 % p.2 ↗ | ssNMR |
| Data acquisition time | ~1 hour p.2 ↗ | pXRD |
| Data acquisition time | 6–24 hours p.2 ↗ | ssNMR |
Operating Limits
- Limit of detection (LOD): 0.9 % w/w p.2 ↗
- Crystalline content analyzed (max): 9.4 % w/w p.2 ↗