US EPA Method 8260 With the Tekmar Lumin and the Agilent 7890B

Technical Reference Summary

Covered Models & Document Scope

Application note for the Teledyne Tekmar Lumin purge-and-trap concentrator with AQUATek 100 autosampler, coupled to an Agilent 7890B GC/MS, demonstrated with US EPA Method 8260 in conjunction with US EPA Method 5030; covers Lumin and AQUATek 100 instrument conditions and VOC calibration/MDL data.

Technical Data

Parameter Value Condition
Valve oven temperature 150 °C p.2 ↗ Lumin, standby
Transfer line temperature 150 °C p.2 ↗ Lumin, standby
Sample mount temperature 90 °C p.2 ↗ Lumin, standby
Standby flow 20 mL/min p.2 ↗ Lumin, standby
Purge ready temperature 35 °C p.2 ↗ Lumin, standby
MCS purge temperature 20 °C p.2 ↗ Lumin, standby
Purge temperature 20 °C p.2 ↗ Lumin, purge
Purge time 11.00 min p.2 ↗ Lumin, purge
Purge flow 40 mL/min p.2 ↗ Lumin, purge
Dry purge temperature 20 °C p.2 ↗ Lumin, purge
Dry purge time No Dry Purge, 0 min p.2 ↗ Lumin, purge
Desorb preheat temperature 245 °C p.2 ↗ Lumin, desorb
Desorb time 2.00 min p.2 ↗ Lumin, desorb
Drain flow 300 mL/min p.2 ↗ Lumin, desorb
Bake time 2.00 min p.2 ↗ Lumin, bake
Bake temperature 280 °C p.2 ↗ Lumin, bake
Condenser bake temperature 180 °C p.2 ↗ Lumin, bake
Bake flow 200 mL/min p.2 ↗ Lumin, bake
Sample loop time 0.35 min p.2 ↗ AQUATek 100
Sample transfer time 0.35 min p.2 ↗ AQUATek 100
Rinse loop time 0.30 min p.2 ↗ AQUATek 100
Sweep needle time 0.30 min p.2 ↗ AQUATek 100
Presweep time 0.25 min p.2 ↗ AQUATek 100
Water temperature 90 °C p.2 ↗ AQUATek 100

Operating Limits

  • Calibration range (all compounds except ketones and t-butanol): 0.5 to 200 ppb p.2 ↗
  • Calibration range (ketones): 1.25 to 500 ppb p.2 ↗
  • Calibration range (t-butanol): 2.5 to 10,000 ppb p.2 ↗
  • MDL standards prepared: seven 0.5 ppb and seven 2 ppb standards p.2 ↗