US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T

Technical Reference Summary

Covered Models & Document Scope

Application note / study for Teledyne Tekmar Lumin purge-and-trap concentrator with AQUATek 100 autosampler and Agilent 7890B GC/5977A MS, demonstrating US EPA Method 524.2 VOC analysis; covers Lumin and AQUATek 100 operating conditions (Table 1) and GC/MS conditions (Table 2), not a product ordering datasheet.

Technical Data

Parameter Value Condition
Valve oven temperature 150 °C p.2 ↗ Lumin, Table 1
Transfer line temperature 150 °C p.2 ↗ Lumin, Table 1
Sample mount temperature 90 °C p.2 ↗ Lumin, Table 1
Purge ready temperature 35 °C p.2 ↗ Lumin, Table 1
MCS purge temperature 20 °C p.2 ↗ Lumin, Table 1
Standby flow 10 mL/min p.2 ↗ Lumin, Table 1
Purge time 11.00 min p.2 ↗ Lumin, Table 1
Purge flow 40 mL/min p.2 ↗ Lumin, Table 1
Dry purge temperature 20 °C p.2 ↗ Lumin, Table 1
Dry purge time 1.00 min p.2 ↗ Lumin, Table 1
Dry purge flow 100 mL/min p.2 ↗ Lumin, Table 1
Desorb time 250 °C p.2 ↗ Lumin, Table 1 (as printed)
Bake time 2.00 min p.2 ↗ Lumin, Table 1
Bake temperature 280 °C p.2 ↗ Lumin, Table 1
MCS bake temperature 200 °C p.2 ↗ Lumin, Table 1
Bake flow 200 mL/min p.2 ↗ Lumin, Table 1
Sample loop time 0.35 min p.2 ↗ AQUATek 100, Table 1
Sample transfer time 0.35 min p.2 ↗ AQUATek 100, Table 1
Rinse loop time 0.30 min p.2 ↗ AQUATek 100, Table 1
Sweep needle time 0.30 min p.2 ↗ AQUATek 100, Table 1
Presweep time 0.25 min p.2 ↗ AQUATek 100, Table 1
Water temperature 90 °C p.2 ↗ AQUATek 100, Table 1
Bake rinse drain cycles 1 p.2 ↗ AQUATek 100, Table 1
Bake rinse drain time 0.35 min p.2 ↗ AQUATek 100, Table 1
Calibration range 0.5 to 50 ppb p.2 ↗ all compounds except ketones and t-butanol

Operating Limits

  • Calibration range (ketones): 1.25 to 125 ppb p.2 ↗
  • Calibration range (t-butanol): 2.5 to 250 ppb p.2 ↗
  • Number of 0.5 ppb standards for MDL: 7 p.2 ↗