Induztr
Engineering Digest Document Viewer Related Models

Stratagene QPCR Mouse Reference Total RNA

Technical Reference Summary

Covered Models & Document Scope

Stratagene QPCR Mouse Reference Total RNA, a high-quality control for quantitative PCR gene-expression analysis composed of total RNA from 11 mouse cell lines; document covers materials provided, storage, assay optimization/standardization/validation protocols and RT-PCR programs.

Technical Data

Parameter Value Condition
Composition total RNA from 11 mouse cell lines p.5 ↗
Storage conditions as stated in Storage Conditions section p.3 ↗
SYBR Green I dye primer optimization range 50–300 nM p.7 ↗
TaqMan probe optimization range 50–600 nM p.7 ↗
Molecular beacon optimization range 200–600 nM p.7 ↗
Passive reference dye dilution 1:500 p.10 ↗ Mx3005P and Mx4000 instruments
Passive reference dye concentration 30 nM p.10 ↗ Mx3000P, Mx3005P, Mx4000 instruments
RT-PCR two-step program initial step 30 minutes at 50°C p.11 ↗ TaqMan probes
RT-PCR two-step program denaturation 10 minutes at 95°C p.11 ↗ TaqMan probes
RT-PCR two-step program cycling 40 cycles of 15 seconds at 95°C p.11 ↗ TaqMan probes
RT-PCR three-step program initial step 30 minutes at 50°C p.11 ↗ Molecular beacons
RT-PCR three-step program denaturation 10 minutes at 95°C p.11 ↗ Molecular beacons
RT-PCR three-step program cycling 40 cycles of 30 seconds at 95°C p.11 ↗ Molecular beacons
Standard curve dilution series 100, 10, 1, 0.1, 0.01 ng p.13 ↗ High abundance target mRNA, 10-fold dilution
Reference RNA volume added per reaction 2.5 μl p.13 ↗ standard curve reactions

Operating Limits

  • SYBR Green I dye primer optimization range: 50–300 nM p.7 ↗
  • TaqMan probe optimization range: 50–600 nM p.7 ↗
  • Molecular beacon optimization range: 200–600 nM p.7 ↗
  • Passive reference dye concentration range: 30–300 nM p.10 ↗