Determination of Multiple Attributes of Monoclonal Antibodies
Technical Reference Summary
Covered Models & Document Scope
Application note describing a fully automated 3D-LC/MS multi-attribute analyzer for monoclonal antibodies: Agilent InfinityLab 2D-LC Solution with 2D multimethod option (SEC, CEX, or HIC in the second dimension) and Agilent 6530 LC/Q-TOF for desalting SEC-MS in the third dimension. Covers trastuzumab originator and trastuzumab-producing CHO clones (clone 3, 6, 8, 10).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| First-dimension column | Agilent Bio-Monolith protein A column, 5 p.4 ↗ | 1D Protein A |
| First-dimension flow rate | 0.75 mL/min p.4 ↗ | 1D Protein A |
| First-dimension mobile phase | 150 mM sodium phosphate pH 7.0 p.4 ↗ | 1D Protein A |
| First-dimension DAD detection | 214 and 280 nm p.4 ↗ | 1D Protein A |
| Second-dimension column | Agilent Bio MAb NP5 PK, 2.1 × 250 mm, 5 p.5 ↗ | 2D CEX |
| Second-dimension column temperature | 25 °C p.5 ↗ | 2D CEX |
| Second-dimension mobile phase A | 25 mM sodium phosphate pH 7.0 p.5 ↗ | 2D CEX |
| Second-dimension flow rate | 0.2 mL/min p.5 ↗ | 2D CEX |
| Second-dimension DAD detection | 280 nm p.5 ↗ | 2D CEX |
| Second-dimension peak width | >0.05 min (5 Hz) p.5 ↗ | 2D CEX |
| Third-dimension column temperature | 25 °C p.6 ↗ | 3D SEC-MS |
| Drying gas flow | 8 L/min p.6 ↗ | 3D SEC-MS |
| Nebulizer pressure | 35 psi p.6 ↗ | 3D SEC-MS |
| Sheath gas flow | 11 L/min p.6 ↗ | 3D SEC-MS |
| Data acquisition range | 1 spectrum/s p.6 ↗ | 3D SEC-MS |
| Calibration concentration range | 0.02–2.0 µg/µL p.8 ↗ | 1D Protein A titer calibration |
| Calibration curve linearity | R² = 0.9998 p.8 ↗ | 1D Protein A titer calibration |
| Sample load for qualitative MS | 100 µg p.11 ↗ | 2D HIC-MS |
Operating Limits
- Second-dimension column temperature: 25 °C p.5 ↗
- Third-dimension column temperature: 25 °C p.6 ↗
- Nebulizer pressure: 35 psi p.6 ↗