Analysis of Cannabinoids and their Metabolites in Human Urine Using the Agilent Chem Elut S Plate by LC/MS/MS

Technical Reference Summary

Covered Models & Document Scope

Application note for the Agilent Chem Elut S 96-well plate (2 mL, square well, full skirt) used for SLE of THC, THC-OH, THC-COOH, CBD and CBN from human urine with LC/MS/MS; includes method parameters, calibration, QC and verification data.

Technical Data

Parameter Value Condition
Plate format 96-well p.2 ↗ Chem Elut S
Well volume 2 mL p.2 ↗ Chem Elut S
Well shape square p.2 ↗ Chem Elut S
Plate skirt full skirt p.2 ↗ Chem Elut S
Sorbent type synthetic sorbent p.2 ↗ Chem Elut S
LC flow rate 0.4 mL/min p.3 ↗
Needle wash 1:1:1:1 ACN/MeOH/IPA/water, 0.2% formic acid p.3 ↗
Ionization mode Electrospray ionization (ESI) p.3 ↗
Nebulizer pressure 22 psi p.3 ↗
Nozzle voltage 0 V p.3 ↗ both positive and negative
Calibration range 2 to 1,000 ng/mL p.9 ↗ human urine matrix
Linearity (R²) >0.99 p.9 ↗ all analytes
LLOQ 2 ng/mL p.8 ↗
Urine volume per sample 50 µL p.4 ↗ calibration standards and QC
IPA pre-addition volume 50 µL p.4 ↗ calibration standards and QC
THC-COOH glucuronide spiking level 75.75 ng/mL p.9 ↗ corresponds to THC-COOH 50 ng/mL
Extraction solvent mixture 9:1 hexane/EtOAc p.7 ↗ elution
Recommended maximum elution volume 15 % of total p.6 ↗ not recommended to use more

Operating Limits

  • Maximum recommended elution volume: 15% of total p.6 ↗
  • Calibration dynamic range: 2 to 1,000 ng/mL p.9 ↗
  • LLOQ: 2 ng/mL p.8 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
400 µL (part number 5610-2004) p.3 ↗ 5610-2004 Sample preparation equipment/consumable referenced in Experimental section
part number 5191-4116 p.3 ↗ 5191-4116 Sample preparation equipment/consumable referenced in Experimental section