| Oven temperature |
70 °C
p.3 ↗ |
8890 GC, 8697-XL method |
| Inlet temperature |
150 °C
p.3 ↗ |
SSL inlet |
| Loop temperature |
80 °C
p.3 ↗ |
8697-XL headspace sampler |
| Transfer line temperature |
90 °C
p.3 ↗ |
heat-traced transfer line |
| Carrier gas |
Helium
p.3 ↗ |
|
| Split ratio |
10:1
p.3 ↗ |
SSL inlet |
| Vial equilibration |
7 min
p.3 ↗ |
8697-XL headspace sampler |
| Control mode |
Constant pressure
p.3 ↗ |
carrier gas control |
| Injection duration |
1 min
p.3 ↗ |
8697-XL headspace sampler |
| Inlet pressure |
21 psi
p.3 ↗ |
SSL inlet |
| GC cycle time |
6 min
p.3 ↗ |
|
| Septum purge |
3 mL/min
p.3 ↗ |
SSL inlet |
| Vial size |
20 mL
p.3 ↗ |
headspace vial |
| Gas saver |
20 mL/min after 2 min
p.3 ↗ |
SSL inlet |
| Vial shaking |
250 shakes/min
p.3 ↗ |
8697-XL headspace sampler |
| Oven program |
40 °C isothermal for
p.3 ↗ |
8890 GC |
| Vial fill mode |
Pressure
p.3 ↗ |
8697-XL headspace sampler |
| FID A/B temperature |
250 °C
p.3 ↗ |
dual FID |
| Fill pressure |
15 psi
p.3 ↗ |
8697-XL headspace sampler |
| FID A/B air flow |
400 mL/min
p.3 ↗ |
dual FID |
| Pressurization gas |
Nitrogen
p.3 ↗ |
8697-XL headspace sampler |
| FID A/B hydrogen flow |
30 mL/min
p.3 ↗ |
dual FID |
| Post-injection purge |
Custom
p.3 ↗ |
8697-XL headspace sampler |
| Headspace sample volume |
500 µL
p.2 ↗ |
450 µL ISTD + 50 µL sample |
| Ethanol linearity range |
0.025 to 0.3 %
p.1 ↗ |
8697-XL with 8890 GC dual-FID |