Application note covering BAC analysis using the 120-vial Agilent 8697 Headspace Sampler -XL Tray coupled to an Agilent 8890 GC System with SSL inlet and dual FID; includes method parameters, consumables (5190-9765, G3440-85035), and retention-time data for DB-BAC1 UI and DB-BAC2 UI columns.

Technical Data

Parameter Value Condition
Oven temperature 70 °C p.3 ↗ 8890 GC, 8697-XL method
Inlet temperature 150 °C p.3 ↗ SSL inlet
Loop temperature 80 °C p.3 ↗ 8697-XL headspace sampler
Transfer line temperature 90 °C p.3 ↗ heat-traced transfer line
Carrier gas Helium p.3 ↗
Split ratio 10:1 p.3 ↗ SSL inlet
Vial equilibration 7 min p.3 ↗ 8697-XL headspace sampler
Control mode Constant pressure p.3 ↗ carrier gas control
Injection duration 1 min p.3 ↗ 8697-XL headspace sampler
Inlet pressure 21 psi p.3 ↗ SSL inlet
GC cycle time 6 min p.3 ↗
Septum purge 3 mL/min p.3 ↗ SSL inlet
Vial size 20 mL p.3 ↗ headspace vial
Gas saver 20 mL/min after 2 min p.3 ↗ SSL inlet
Vial shaking 250 shakes/min p.3 ↗ 8697-XL headspace sampler
Oven program 40 °C isothermal for p.3 ↗ 8890 GC
Vial fill mode Pressure p.3 ↗ 8697-XL headspace sampler
FID A/B temperature 250 °C p.3 ↗ dual FID
Fill pressure 15 psi p.3 ↗ 8697-XL headspace sampler
FID A/B air flow 400 mL/min p.3 ↗ dual FID
Pressurization gas Nitrogen p.3 ↗ 8697-XL headspace sampler
FID A/B hydrogen flow 30 mL/min p.3 ↗ dual FID
Post-injection purge Custom p.3 ↗ 8697-XL headspace sampler
Headspace sample volume 500 µL p.2 ↗ 450 µL ISTD + 50 µL sample
Ethanol linearity range 0.025 to 0.3 % p.1 ↗ 8697-XL with 8890 GC dual-FID

Operating Limits

  • Ethanol linearity range: 0.025 to 0.3% p.1 ↗
  • Vial-to-vial reproducibility: 2.0% RSD at 0.05% concentration p.1 ↗
  • Ethanol carryover: none at 0.4% concentration p.1 ↗
  • Headspace sampler vial capacity: 120 vials (-XL Tray) p.1 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent Blood Alcohol Checkout Mix p.2 ↗ 5190-9765 Repeatability/resolution test mix
Agilent Ethanol Calibration Kit p.2 ↗ G3440-85035 Ethanol calibration