| LC/MS/MS injection volume |
2 μL
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS stop time |
10 min
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS gas temperature |
120 °C
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS gas flow |
20 L/min
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS nebulizer pressure |
40 psi
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS sheath gas heater temperature |
225 °C
p.3 ↗ |
LC/MS/MS conditions |
| LC/MS/MS low-pressure RF |
60 V
p.3 ↗ |
positive polarity |
| LC/MS/MS low-pressure RF |
60 V
p.3 ↗ |
negative polarity |
| LC/MS/MS polarity |
Positive and negative
p.3 ↗ |
LC/MS/MS conditions |
| GC/MS/MS carrier gas |
Helium
p.3 ↗ |
GC/MS/MS conditions |
| GC/MS/MS column 1 flow |
1.0 mL/min
p.3 ↗ |
GC/MS/MS conditions |
| GC/MS/MS column 2 flow |
1.4 mL/min
p.3 ↗ |
GC/MS/MS conditions |
| GC/MS/MS injection volume |
1 µL
p.3 ↗ |
cold splitless |
| GC/MS/MS inlet liner |
Agilent inlet liner, Ultra Inert, splitless
p.3 ↗ |
GC/MS/MS conditions |
| Captiva EMR–GPF cartridge volume |
3 mL
p.3 ↗ |
Captiva EMR–GPF |
| Captiva EMR–GPF part number |
5610-2090
p.3 ↗ |
Captiva EMR–GPF cartridge, 3 mL |
| Ceramic homogenizer pack size |
100 pk
p.3 ↗ |
50 mL tubes, part number 5982-9313 |
| Spiking level for recovery study |
10 ng/g
p.13 ↗ |
230 targets |
| Recovery acceptance criterion |
60 to 120 %
p.16 ↗ |
accuracy |
| RSD acceptance criterion |
<20 %
p.16 ↗ |
reproducibility |
| Matrix effect acceptance criterion (LC/MS/MS) |
70 to 100 %
p.16 ↗ |
LC/TQ only |
| Calibration curve acceptance criterion |
R2 >0.99
p.16 ↗ |
calibration curve |
| Number of pesticides targeted |
240
p.1 ↗ |
multiresidue method |
| Number of LC-amenable pesticides |
129
p.2 ↗ |
LC/MS/MS |
| Number of pesticides in quantitative analysis |
230
p.10 ↗ |
bell pepper, LC/MS/MS and GC/MS/MS |