Using a Deans Switch to Divert Derivatizing Agents from the EI Source for Improved Robustness
Technical Reference Summary
Covered Models & Document Scope
Application note on using a Deans switch to divert derivatization agents (MSTFA) from the EI source in anti-doping GC/MS/MS workflows. Covers two instrument configurations: Agilent 7000 Series GC/TQ with hydrogen carrier gas and Agilent 7010 Series GC/TQ with helium carrier gas; two inlet types (split/splitless inlet SSL and programmable temperature vaporizer inlet).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Purge flow to split vent | 50 mL/min p.6 ↗ | SSL inlet, at 0.75 min |
| Purge flow to split vent | 50 mL/min p.6 ↗ | PTV inlet, at 0.75 min |
| Injection volume | 1.5 µL p.6 ↗ | SSL inlet |
| Switching time | 4.9 min p.6 ↗ | SSL inlet, from vent to MS |
| Column temperature program | 2.5 °C/min to 199 °C (0 min hold) °C p.6 ↗ | SSL inlet |
| Carrier gas and flow rate | Hydrogen, 1 mL/min constant flow p.6 ↗ | SSL inlet |
| Transfer line temperature | 300 °C p.6 ↗ | SSL inlet |
| Electron energy | 70 eV p.6 ↗ | SSL inlet |
| Electron energy | 70 eV p.6 ↗ | PTV inlet |
| Quench gas (helium) | 2.25 mL/min p.6 ↗ | PTV inlet |
| Collision gas (nitrogen) | 1.5 mL/min p.6 ↗ | SSL inlet |
| Collision gas (nitrogen) | 1.5 mL/min p.6 ↗ | PTV inlet |
| Ion source temperature | 280 °C p.6 ↗ | SSL inlet |
| Ion source temperature | 280 °C p.6 ↗ | PTV inlet |
| Tune | Etune: atunes.eiex.tune.xml p.6 ↗ | SSL inlet |
| Cleaning interval | 350–400 to ≥ 1,500 injections p.1 ↗ | Agilent 7000 Series GC/TQ, hydrogen carrier gas |
Operating Limits
- Source cleaning interval improvement (7000 Series GC/TQ, hydrogen): from approximately 350–400 to ≥ 1,500 injections p.1 ↗
- Source cleaning interval improvement (7010 Series GC/TQ, helium): from approximately every 300 to 350 injections to approximately 900 injections p.1 ↗
- Routine batch size: 80 samples p.1 ↗
- Source cleaning frequency with routine batches: shifted from weekly to every 2 to 3 weeks p.1 ↗