Dynamic MRM/Scan Mode, Sensitive Quantitation in Complex Foods by GC/TQ

Technical Reference Summary

Covered Models & Document Scope

Application note covering simultaneous dMRM/scan data acquisition on Agilent 8890/7000E and 8890/7010C triple quadrupole GC/MS (GC/TQ) systems for pesticide residue analysis in spinach, walnut and cayenne pepper matrices; no ordering codes or product variants are defined.

Technical Data

Parameter Value Condition
GC system Agilent 8890 with fast oven, auto injector p.3 ↗ 8890/7000E and 8890/7010C
Mass spectrometer 7000E / 7010C triple quadrupole p.2 ↗
Vacuum pump Performance turbo p.3 ↗
Source temperature 280 °C p.3 ↗
Acquisition mode Simultaneous dMRM/scan p.3 ↗
Initial oven temperature 60 °C p.3 ↗
Initial oven hold 1 min p.3 ↗
Final temperature 1 170 °C p.3 ↗ ramp 1
Ramp rate 2 10 °C/min p.3 ↗
Final temperature 2 310 °C p.3 ↗ ramp 2
Final hold 2 2.25 min p.3 ↗ ramp 2
Total run time 20 min p.3 ↗
Post run time 1.5 min p.3 ↗
Equilibration time 0.25 min p.3 ↗
Post run inlet temperature 310 °C p.3 ↗
Post run total flow 25 mL/min p.3 ↗
Column 1 Agilent HP-5ms UI, 15 m × 0.25 mm, 0.25 µm p.3 ↗
He quench gas flow 2.25 mL/min p.3 ↗
N2 collision gas flow 1.5 mL/min p.3 ↗
Total MRMs (dMRM mode) 614 p.3 ↗
Minimum dwell time 6.85 ms p.3 ↗
Minimum cycle time 69.8 ms p.3 ↗
Scan type MS1 scan p.3 ↗ full scan
Scan range 45 to 450 m/z p.3 ↗ full scan

Operating Limits

  • Calibration range: 0.1 to 5,000 ppb p.11 ↗
  • Minimum dwell time: 6.85 ms p.3 ↗
  • Minimum cycle time: 69.8 ms p.3 ↗