Analysis of Essential Oils Using Comprehensive Two-Dimensional Gas Chromatography (GC × GC)

Technical Reference Summary

Covered Models & Document Scope

Application note describing a GC × GC workflow for the analysis of essential oils using a flow modulator (RFM) combined with a high-resolution accurate-mass GC/Q-TOF system and FID detection; no product models, sizes or ordering codes are covered.

Technical Data

Parameter Value Condition
Injection mode Split; 250:1 p.3 ↗
Injection volume 0.5 µL p.3 ↗
Inlet temperature 300 °C p.3 ↗
Electron energy 70 eV p.3 ↗
Emission current 5 µA p.3 ↗
FID temperature 300 °C p.3 ↗
FID H2 flow 30 mL/min p.3 ↗
FID air flow 400 mL/min p.3 ↗
MS carrier gas flow range 1 to 2 mL/min p.2 ↗
Scanning mass analyzer maximum acquisition rate near 20 Hz p.2 ↗

Operating Limits

  • MS carrier gas flow constraint: 1 to 2 mL/min p.2 ↗
  • Scanning mass analyzer maximum acquisition rate: near 20 Hz p.2 ↗