Analyse gasförmiger Produkte aus Technologien zur

Technical Reference Summary

Covered Models & Document Scope

Application Note (Agilent Technologies) describing a multidimensional GC configuration with TCD, FID and nickel catalyst for analysis of gaseous products from CO2 utilization technologies; covers H2, O2, N2, CH4, CO, CO2 and C1–C6 hydrocarbons; document is an application note, not a product datasheet, and lists no instrument model/size variants or ordering codes.

Technical Data

Parameter Value Condition
Detection range, lower limit 0.1 ppm p.1 ↗ overall system, TCD + FID with catalyst
Detection range, upper limit 100 % p.1 ↗ overall system, TCD + FID with catalyst
TCD detection limit 100 ppm p.1 ↗ H2, O2, N2, CH4, CO, CO2
FID with catalyst detection limit 0.1 ppm p.1 ↗ CH4, CO, CO2, C2–C6 hydrocarbons
Analytes H2, O2, N2, CH4, CO, CO2, C1–C6 hydrocarbons p.1 ↗
Detectors TCD and FID p.1 ↗ multidimensional configuration
Catalyst Nickel catalyst p.1 ↗ positioned at FID
Valve 1 switching ON/OFF position p.3 ↗ used to divert compounds with more than 2 carbon atoms (C2+)
Pressure controller three-channel pressure controller p.3 ↗

Operating Limits

  • Detection range of the method: 0.1 ppm to 100 % p.1 ↗
  • TCD detection limit: 100 ppm p.1 ↗
  • FID with catalyst detection limit: 0.1 ppm p.1 ↗