Group-Type Analysis of Jet Fuel Using the Agilent Flow-Modulated GC×GC-FID

Technical Reference Summary

Covered Models & Document Scope

Application note for group-type analysis of jet fuel according to NB/SH/T 6078-2023 using the Agilent 8890 reverse fill/flush flow-modulated GC×GC-FID system; covers three column sets (Column Set 1, Column Set 2, Column Set 3) with specified primary, secondary, and monitor columns.

Technical Data

Parameter Value Condition
Detector temperature 300 °C p.3 ↗ Column Set 1 and 2
Hydrogen flow 35 mL/min p.3 ↗ Column Set 1 and 2
Air flow 350 mL/min p.3 ↗ Column Set 1 and 2
Acquisition rate 200 Hz p.3 ↗ Column Set 1 and 2
Primary column Agilent DB-1, 20 m × 0.18 mm, 0.4 μm (p/n 121-1023) p.3 ↗ Column Set 1
Primary column flow rate 0.5 mL/min p.3 ↗ Column Set 1
Secondary column p/n 123-7157 p.3 ↗ Column Set 1
Secondary column flow rate 25 mL/min p.3 ↗ Column Set 1
Monitor column Deactivated fused silica, 1.9 m × 0.1 mm (p/n 160-2635-10) p.3 ↗ Column Set 1
Monitor column flow rate 0.63 mL/min p.3 ↗ Column Set 1
Carrier gas He p.3 ↗ Column Set 1
Carrier gas He p.3 ↗ Column Set 2

Operating Limits

  • Retention time repeatability (%RSD of 1RT): ~0 p.1 ↗
  • Retention time repeatability (%RSD of 2RT): ≤ 0.27 p.1 ↗
  • Quantitative precision (%RSD) for all nontrace compounds across 10 replicates: < 2.0 p.1 ↗
  • Linearity (R²) across eight concentration levels: > 0.999 p.1 ↗