Agilent LUMA Multichannel Vacuum Ultraviolet Detector on Agilent 8890 GC
Technical Reference Summary
Covered Models & Document Scope
Application note for the LUMA multichannel vacuum ultraviolet (VUV) detector used with an Agilent 8890 GC and J&W DB-1 column (30 m × 320 µm, 1 µm); covers detector evaluation with the Polar ISO Column Text Mix standard, 12-band configuration, and eight analytes at 7.8–250 ppm.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection volume | 1 µL p.2 ↗ | |
| Inlet temperature | 250 °C p.2 ↗ | Split/Splitless, 5:1 split |
| Oven program initial temperature | 40 °C p.2 ↗ | hold for 1 min |
| Oven program ramp rate | 20 °C/min p.2 ↗ | to 260 °C |
| Oven program final temperature | 260 °C p.2 ↗ | |
| Column flow | 6.5 mL/min p.2 ↗ | hydrogen carrier |
| LUMA temperature setpoint | 275 °C p.2 ↗ | |
| Number of detector bands | 12 p.2 ↗ | simultaneously acquired |
| Standard concentration range | 7.8 to 250 ppm p.3 ↗ | six standards |
| Average peak area RSD | < 2 % p.3 ↗ | five consecutive injections, 7.8 and 250 ppm |
| Average R² | 0.999 p.3 ↗ | eight compounds |
| S/N for Aniline | 415 p.3 ↗ | 7.8 ppm, Band 7 |
| S/N for 2-Chlorophenol | 398 p.3 ↗ | 7.8 ppm, Band 7 |
| S/N for 1-Octanol | 155 p.3 ↗ | 7.8 ppm, Band 2 |
| S/N for 2-Nonanone | 169 p.3 ↗ | 7.8 ppm, Band 2 |
| S/N for 2-Dodecanol | 152 p.3 ↗ | 7.8 ppm, Band 2 |
| S/N for Methyl Laurate (C12:0) | 150 p.3 ↗ | 7.8 ppm, Band 2 |
| S/N for n-Heptadecane (C17) | 170 p.3 ↗ | 7.8 ppm, Band 2 |
| S/N for n-Nonadecane (C19) | 158 p.3 ↗ | 7.8 ppm, Band 2 |
| Peak symmetry for Aniline | 0.95 p.3 ↗ | Band 7 |
| Peak symmetry for 2-Chlorophenol | 0.98 p.3 ↗ | Band 7 |
| Peak symmetry for 1-Octanol | 0.97 p.3 ↗ | Band 2 |
| Peak symmetry for 2-Nonanone | 0.98 p.3 ↗ | Band 2 |
| Peak symmetry for 2-Dodecanol | 0.94 p.3 ↗ | Band 2 |
| Peak symmetry for Methyl Laurate (C12:0) | 1.00 p.3 ↗ | Band 2 |
Operating Limits
- LUMA detector temperature setpoint: 275 °C p.2 ↗
- GC inlet temperature: 250 °C p.2 ↗
- Oven temperature range: 40 °C to 260 °C p.2 ↗
- Band 2 wavelength range: 130 to 143 nm p.3 ↗
- Band 7 wavelength range: 189 to 193 nm p.3 ↗
- Band 1 wavelength range: 118 to 130 nm p.3 ↗