Determination of Ultratrace Impurities in Semiconductor Photoresist Using ICP-MS/MS
Technical Reference Summary
Covered Models & Document Scope
Application note covering the Agilent 8900 #200 ICP-MS/MS (ICP-QQQ) configured with a 200 µL/min MicroFlow PFA nebulizer and O2/Ar option gas, used for determination of 20 elemental contaminants in IC photoresist after simple dilution in PGMEA solvent; tune modes NH3, H2 and O2 are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Nebulizer flow rate | 200 µL/min p.3 ↗ | 8900 #200, MicroFlow PFA nebulizer |
| Option gas | 100% O2 p.3 ↗ | O2/Ar mix |
| RF power | 1000 W p.4 ↗ | NH3 tune mode |
| Sampling depth | 18.0 mm p.4 ↗ | NH3 tune mode |
| Makeup gas | 0.60 L/min p.4 ↗ | NH3 tune mode |
| O2/Ar option gas | 0.25 L/min p.4 ↗ | NH3 tune mode |
| Omega lens | 5 V p.4 ↗ | NH3 tune mode |
| Axial acceleration | 0.5 V p.4 ↗ | NH3 tune mode |
Operating Limits
- Maximum acceptable metallic impurity level in photoresist: 1–10 ppb p.6 ↗