Utilisation du PSD pour le rétrobalayage du GC Agilent 8890
Technical Reference Summary
Covered Models & Document Scope
Agilent 8890 GC with PSD (pneumatic switching device) EPC module and Deans switch backflush configuration; application note, no product variants or ordering codes.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Purge flow range | 3 to 30 mL/min p.2 ↗ | PSD purge flow setting |
| Purge flow default | 3 mL/min p.2 ↗ | PSD purge flow setting |
| PSD pressure | 70 psi p.3 ↗ | 4.5 mL/min column 2/3 |
| Inlet type | MMI p.3 ↗ | |
| MMI program | 100 °C (0.02 minute), 900 °C/min jusqu'à p.3 ↗ | |
| Column | DB-HT Sim Dis Agilent J&W, 5 m × 530 µm p.3 ↗ | |
| Second column | DB-17ht Agilent J&W (réf. 122-1831) p.3 ↗ | |
| Syringe | Gamme bleue, 5 µL, conique p.3 ↗ | |
| Inlet insert | Ultra-inerte, division, laine de verre p.3 ↗ | |
| Carrier gas | hélium p.2 ↗ | constant flow mode |
| Deans switch isolation window | 20.1 to 20.4 min p.2 ↗ |
Operating Limits
- PSD purge flow setting range: 3 to 30 mL/min p.2 ↗
- PSD purge flow default: 3 mL/min p.2 ↗