Analytical Development of a Four-Stream Multiplexed LC/MS Method for the Simultaneous Determination of SDMA and ADMA
Technical Reference Summary
Covered Models & Document Scope
Application note describing an LC/MS/MS method for simultaneous determination of ADMA and SDMA on a four-stream Agilent StreamSelect LC/MS system with an Agilent 6490 triple quadrupole LC/MS; no product model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analytical column | Phenomenex Kinetix 2.6 μM 2.1 × 50 mm, alternatively, an Agilent Poroshell p.2 ↗ | |
| Column temperature | Ambient p.2 ↗ | |
| Injection volume | 3 μL p.2 ↗ | |
| Mobile phase A | H2O + 0.005 % HFBA p.2 ↗ | |
| Mobile phase B | Methanol + 0.005 % HFBA p.2 ↗ | |
| Flow rate | 0.4 mL minutes p.2 ↗ | |
| Stop time | 5 minutes p.2 ↗ | |
| Isocratic separation | 5 % mobile phase B for 2.5 minutes with a wash at 95 %B for 2.5 minutes p.2 ↗ | |
| Instrument | Agilent 6490 triple quadrupole LC/MS p.2 ↗ | |
| Ion mode | Positive p.2 ↗ | |
| Drying gas temperature | 250 °C p.2 ↗ | |
| Gas flow | 14 L/min p.2 ↗ | |
| Nebulizer | 20 psi p.2 ↗ | |
| Sheath gas temperature | 300 °C p.2 ↗ | |
| Sheath gas flow | 12 p.2 ↗ | |
| Capillary voltage | 3,000 p.2 ↗ | |
| EMV | +100 V p.2 ↗ | |
| Intra-assay imprecision ADMA | <1.7 % p.2 ↗ | |
| Intra-assay imprecision SDMA | <2.4 % p.2 ↗ | |
| Inter-assay imprecision ADMA | <3.0 % p.2 ↗ | |
| Inter-assay imprecision SDMA | <2.8 % p.2 ↗ | |
| Mean bias at 30 ng/mL ADMA | 0.7 % p.2 ↗ | |
| Mean bias at 30 ng/mL SDMA | 2.5 % p.2 ↗ | |
| Mean bias at 90 ng/mL ADMA | 1.2 % p.2 ↗ | |
| Mean bias at 90 ng/mL SDMA | 3.9 % p.2 ↗ |
Operating Limits
- Calibrator concentration range: 10–500 ng/mL p.2 ↗
- Column temperature: Ambient p.2 ↗
- Drying gas temperature: 250 °C p.2 ↗
- Sheath gas temperature: 300 °C p.2 ↗