Analysis of Cannabinoids and Their Metabolites in Urine Using the
Technical Reference Summary
Covered Models & Document Scope
Application note for the analysis of six cannabinoids (CBD, CBDA, CBN, THC, THC-COOH, THC-OH) and their deuterated internal standards in urine using the Agilent MassHunter StreamSelect LC/MS system with an Agilent 6470 triple quadrupole MS and four Agilent 1290 Infinity II UHPLC streams; no product model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analytical column | Agilent Poroshell 120 EC-C18, 2.1 × 50 mm, 2.7 µm (p/n 699775-902T) p.3 ↗ | |
| Column temperature | 50 °C p.3 ↗ | |
| Injection volume | 5 µL p.3 ↗ | |
| Autosampler temperature | 4 °C p.3 ↗ | |
| Needle wash | 50 % Methanol p.3 ↗ | |
| Mobile phase A | Milli-Q water with 5 p.3 ↗ | |
| Mobile phase B | Methanol with 0.01 % formic acid p.3 ↗ | |
| Sheath flow | 12 L/min p.3 ↗ | |
| Nebulizer pressure | 45 psi p.3 ↗ | |
| Capillary voltage | 500 V p.3 ↗ | |
| Q1/Q3 resolution | 0.7 unit p.3 ↗ | |
| Dwell time | 18 ms p.3 ↗ | |
| Fragmentor voltage | 200 V p.3 ↗ | |
| Calibrator range | 5 to 5,000 ng/mL p.3 ↗ | |
| Chromatographic method duration | six-minute p.1 ↗ | |
| Calibration linearity (R²) | >0.996 p.1 ↗ | |
| Retention time reproducibility (%RSD) | 0.43 to 1.54 % p.1 ↗ | n = 1,000 |
| Quantitation accuracy | 95.3 to 109.1 % p.1 ↗ | 200 ng/mL sample |
Operating Limits
- Calibrator concentration range: 5 to 5,000 ng/mL p.3 ↗
- Column temperature: 50 °C p.3 ↗
- Autosampler temperature: 4 °C p.3 ↗