ASTM D8351-22: Elemental Analysis of Biodiesel using MP-AES
Technical Reference Summary
Covered Models & Document Scope
Application study of biodiesel elemental analysis (Ca, Mg, P, K, Na) per ASTM D8351-22 using the Agilent 4210 MP-AES with organics kit (EGCM, OneNeb Series 2 nebulizer, double-pass glass cyclonic spray chamber), SPS 4 autosampler and optional Agilent 4107 nitrogen generator; results from five laboratories.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Read time | 3 to 10 s p.2 ↗ | 4210 MP-AES operating conditions |
| Number of replicates | 3 p.2 ↗ | 4210 MP-AES operating conditions |
| Stabilization time | 30 s p.2 ↗ | 4210 MP-AES operating conditions |
| Rinse time | Approximately 60 s p.2 ↗ | depends on lab |
| Nebulizer flow rate | 0.5 p.3 ↗ | all elements, organics kit |
| EGCM injection | High p.3 ↗ | Ca 396.847 nm |
| EGCM injection | Medium p.3 ↗ | Mg 285.213 nm |
| EGCM injection | Medium p.3 ↗ | P 213.618 nm |
| EGCM injection | High p.3 ↗ | K 766.491 nm |
| EGCM injection | High p.3 ↗ | Na 588.995 nm |
| Background correction | Auto p.3 ↗ | all elements |
| Wavelength | 396.847 nm p.3 ↗ | Calcium |
| Wavelength | 285.213 nm p.3 ↗ | Magnesium |
| Wavelength | 213.618 nm p.3 ↗ | Phosphorus |
| Wavelength | 766.491 nm p.3 ↗ | Potassium |
| Wavelength | 588.995 nm p.3 ↗ | Sodium |
| Method detection limit (MDL) | 0.01 ppm p.4 ↗ | Calcium 396.847 nm |
| Method detection limit (MDL) | 0.02 ppm p.4 ↗ | Magnesium 285.213 nm |
| Method detection limit (MDL) | 0.19 ppm p.4 ↗ | Phosphorus 213.618 nm |
| Method detection limit (MDL) | 0.04 ppm p.4 ↗ | Potassium 766.491 nm |
| Method detection limit (MDL) | 0.02 ppm p.4 ↗ | Sodium 588.995 nm |
Operating Limits
- Method detection limit, Calcium: 0.01 ppm p.4 ↗
- Method detection limit, Magnesium: 0.02 ppm p.4 ↗
- Method detection limit, Phosphorus: 0.19 ppm p.4 ↗
- Method detection limit, Potassium: 0.04 ppm p.4 ↗
- Method detection limit, Sodium: 0.02 ppm p.4 ↗