Induztr
Engineering Digest Document Viewer Related Models

Analyse des dioxines dans l’alimentation humaine et animale

Technical Reference Summary

Covered Models & Document Scope

Application note (Agilent publication 5991-9399FR) on the analysis of dioxins in human and animal feed using a GC-QQQ Intuvo 9000/7010 system; covers two chromatographic methods: a 52.5-minute method and a 31.12-minute accelerated method.

Technical Data

Parameter Value Condition
Injection volume 1 μL p.2 ↗ 52.5-minute method
Column flow 1.6 mL/min p.2 ↗ 52.5-minute method
Column flow 1.6 mL/min p.2 ↗ 31.12-minute method
Guard Chip temperature 310 °C p.2 ↗ 52.5-minute method
Guard Chip temperature 310 °C p.2 ↗ 31.12-minute method
Bus temperature 325 °C p.2 ↗ 52.5-minute method
Bus temperature 325 °C p.2 ↗ 31.12-minute method
Splitless pulsed pressure 60 psi p.2 ↗ 0.6 minute, 50 mL/min
Calibration levels 6 points p.3 ↗ 52.5-minute method
Calibration range (Tetra) 50 to 80 000 fg/µL p.3 ↗ L1–L6
Calibration range (Penta) 100 to 160 000 fg/µL p.3 ↗ L1–L6
Calibration range (Hexa) 100 to 160 000 fg/µL p.3 ↗ L1–L6
Calibration range (Hepta) 200 to 320 000 fg/µL p.3 ↗ L1–L6
Calibration range (Octa) 200 to 320 000 fg/µL p.3 ↗ L1–L6
Coefficient of determination R² (2378-TCDD) 0.99997 p.3 ↗ 52.5-minute method
Coefficient of determination R² (OCDD) 0.99999 p.3 ↗ 52.5-minute method
MDL (2378-TCDF) 3.8184 fg/µL p.4 ↗
LOQ (2378-TCDF) 13.5334 fg/µL p.4 ↗
LOD (2378-TCDF) 4.06 fg/µL p.4 ↗
MDL (OCDD) 16.7276 fg/µL p.4 ↗
LOQ (OCDD) 59.2874 fg/µL p.4 ↗
LOD (OCDD) 17.7862 fg/µL p.4 ↗
Retention time (2378-TCDD) 27.835 min p.4 ↗
Retention time (OCDD) 48.83 min p.4 ↗

Operating Limits

  • Method duration (long method): 52.5 min p.2 ↗
  • Method duration (accelerated method): 31.12 min p.2 ↗
  • Maximum injection volume: 1 μL p.2 ↗
  • Minimum injection volume: 0.5 μL p.2 ↗