Induztr
Engineering Digest Document Viewer Related Models

Osmotic Fragility of Red Blood Cells Quantified on the Agilent NovoCyte Flow Cytometer

Technical Reference Summary

Covered Models & Document Scope

Application note on flow cytometry osmotic fragility testing of red blood cells using the Agilent NovoCyte flow cytometer; covers sample preparation, DI water spiking, and effects of incubation time and temperature. No instrument model variants or ordering codes are described.

Technical Data

Parameter Value Condition
Blood dilution step 1 20 µL blood in 1 mL PBS p.3 ↗
Blood dilution step 2 10 µL diluted blood in 1.1 mL PBS p.3 ↗
Final PBS concentration 55 % PBS p.3 ↗
Baseline sample volume 20 µL p.3 ↗
Flow rate 50 µL/minute p.3 ↗
DI water spike volume 0.9 mL p.3 ↗
Additional acquisition volume 100 µL p.3 ↗
Acquisition segment duration 10 second p.3 ↗
Normal donor residual red cells ~65 % RRC p.3 ↗
Residual red cells after overnight incubation 38 % RRC p.3 ↗ 4 °C
Sample analysis window after collection within two hours p.3 ↗ EDTA tubes

Operating Limits

  • Recommended test timing after blood collection: immediately after blood collection (within two hours) p.3 ↗
  • Incubation temperature range tested: 4 to 37 °C p.3 ↗