Induztr
Engineering Digest Document Viewer Related Models

Analyse von Restlösemitteln gemäß USP-Methode <467> auf dem

Technical Reference Summary

Covered Models & Document Scope

Application Note: analysis of residual solvents according to USP <467> using the Agilent 8890 GC system with 7697A headspace sampler, dual FID, and two columns: Agilent J&W DB-Select 624 UI for 467 (30 m × 0.32 mm, 1.8 µm, part no. 123-0334UI) and Agilent J&W HP-INNOWax.

Technical Data

Parameter Value Condition
Carrier gas Helium, constant flow, 2 ml/min p.3 ↗
Inlet type Split/Splitless p.3 ↗
Oven temperature 85 °C p.3 ↗
Column 1 J&W DB-Select 624 UI für 467, 30 m × 0.32 mm, 1.8 µm p.2 ↗ part no. 123-0334UI
Column 2 J&W HP-INNOWax p.2 ↗
Detector dual flame ionization detectors (FID) p.2 ↗
Headspace sampler Agilent 7697A p.2 ↗
Sample vials clear 10-ml headspace vials with crimp cap p.2 ↗ part no. 5190-2285
Septa Advanced Green, non-stick inlet septa p.2 ↗ part no. 5183-4759-100
Splitter inert T-piece for Capillary Flow Technology p.2 ↗ part no. G3184-60065
Ferrules short graphite ferrules for 0.1–0.32 mm columns, 10/pk p.2 ↗ part no. 5080-8853
Inlet liner 2 mm, splitless, straight, deactivated p.2 ↗ part no. 5181-8818
Headspace transfer line / pre-CFT column deactivated fused silica, 30 m × 0.25 mm ID × 0.35 mm OD p.2 ↗ part no. 160-2255-30

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Sample vials p.2 ↗ 5190-2285 clear 10-ml headspace vials with crimp cap
Septa p.2 ↗ 5183-4759-100 Advanced Green, non-stick inlet septa
Splitter p.2 ↗ G3184-60065 inert T-piece for Capillary Flow Technology
Ferrules p.2 ↗ 5080-8853 short graphite ferrules for 0.1–0.32 mm columns, 10/pk
Ferrules p.2 ↗ G3188-27502 UltiMetal Plus, flexible metal, for 0.32 mm fused silica capillaries, 10/pk
Inlet liner p.2 ↗ 5181-8818 2 mm, splitless, straight, deactivated
Headspace transfer line / pre-CFT column p.2 ↗ 160-2255-30 deactivated fused silica, 30 m × 0.25 mm ID × 0.35 mm OD
Column 1 p.2 ↗ 123-0334UI J&W DB-Select 624 UI für 467, 30 m × 0.32 mm, 1.8 µm