Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model BD S 56

Technical Reference Summary

Covered Models & Document Scope

Model BD-S 56 standard incubator with natural convection, 62 L interior volume. Two variants: BDS056-230V (Art.-No. 9090-0016, 230 V 1~ 50/60 Hz, CEE 7/7 plug) and BDS056UL-120V (Art.-No. 9090-0017, 120 V 1~ 60 Hz, NEMA 5-15 plug).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 70 °C °C p.2 ↗ BDS056-230V
Temperature uniformity at 37°C 0.5 ± K K p.2 ↗ BDS056-230V
Temperature fluctuation at 37°C 0.3 ± K K p.2 ↗ BDS056-230V
Heating up time to 37°C 230 min p.2 ↗ BDS056-230V
Recovery time after door was opened for 30 s at 37°C 15 min p.2 ↗ BDS056-230V
Rated Voltage 230 V p.2 ↗ BDS056-230V
Power frequency 50/60 Hz p.2 ↗ BDS056-230V
Nominal power 0,3 kW p.2 ↗ BDS056-230V
Unit fuse 6,3 A p.2 ↗ BDS056-230V
Phase (Nominal voltage) 1~ p.2 ↗ BDS056-230V
Interior volume 62 L p.2 ↗ BDS056-230V
Net weight of the unit (empty) 35 kg p.2 ↗ BDS056-230V
Load per rack 15 kg p.2 ↗ BDS056-230V
Permitted load 45 kg p.2 ↗ BDS056-230V
Wall clearance back 160 mm p.2 ↗ BDS056-230V
Wall clearance sidewise 100 mm p.2 ↗ BDS056-230V
Width net 614 mm p.2 ↗ BDS056-230V
Height net 625 mm p.2 ↗ BDS056-230V
Depth net 577 mm p.2 ↗ BDS056-230V
Interior width 400 mm p.2 ↗ BDS056-230V
Interior height 440 mm p.2 ↗ BDS056-230V
Interior depth 350 mm p.3 ↗ BDS056-230V
Energy consumption at 37°C 25 Wh/h p.3 ↗ BDS056-230V
Number of shelves (std./max.) 1/3 p.3 ↗ BDS056-230V
Rated Voltage 120 V p.2 ↗ BDS056UL-120V, 60 Hz

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 70 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 45 kg p.2 ↗
  • Load per rack: 15 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ BDS Product line: Solid.Line standard incubator
4-6 p.1 ↗ 056 Model size (56 series, 62 L interior volume)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz supply with NEMA 5-15 plug (absent = standard 230 V variant)
suffix p.1 ↗ 230V Mains voltage variant 230 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ 120V Mains voltage variant 120 V 1~ 60 Hz, NEMA 5-15 plug

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
BDS056-230V p.1 ↗ 9090-0016 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard version
BDS056UL-120V p.1 ↗ 9090-0017 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard UL version

Related models