Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model CB 260

Technical Reference Summary

Covered Models & Document Scope

Model CB 260 CO2 incubator, interior volume 267 L. Four model versions: CB260-230V (standard, 200–230 V), CB260-230V-O (with O2 control 0.2–20%), CB260UL-120V (standard, 100–120 V, NEMA 5-20), CB260UL-120V-O (with O2 control 0.2–20%, 100–120 V).

Technical Data

Parameter Value Condition
Interior volume 267 L p.3 ↗ all models
Rated voltage 200...230 V p.3 ↗ CB260-230V / CB260-230V-O
Rated voltage 100...120 V p.3 ↗ CB260UL-120V / CB260UL-120V-O
Power frequency 50/60 Hz p.3 ↗ all models
Nominal power 1,5 kW p.3 ↗ all models
Unit fuse 10,0 A p.3 ↗ CB260-230V / CB260-230V-O
Unit fuse 16,0 A p.3 ↗ CB260UL-120V / CB260UL-120V-O
Phase (nominal voltage) 1~ p.3 ↗ all models
Temperature range to 50 °C p.3 ↗ all models
Temperature uniformity at 37°C 0.4 ± K p.3 ↗ all models
Temperature fluctuation at 37°C 0.1 ± K p.3 ↗ all models
Recovery time 6 min p.3 ↗ all models
Humidity range 90 ...95 % RH p.3 ↗ all models
CO2 range 0 ...20 Vol.-% CO2 p.3 ↗ all models
CO2 measuring technology IR p.3 ↗ all models
CO2 recovery time 5 min p.3 ↗ all models
Standard O2 control range 0.2 ...20 Vol.-% O2 p.3 ↗ CB260-230V-O / CB260UL-120V-O
Net weight of the unit (empty) 125 kg p.3 ↗ all models
Load per rack 10 kg p.3 ↗ all models
Permitted load 40 kg p.4 ↗ all models
Wall clearance back 100 mm p.4 ↗ all models
Wall clearance sidewise 50 mm p.4 ↗ all models
Width net 740 mm p.4 ↗ all models
Height net 1,020 mm p.4 ↗ all models
Depth net 785 mm p.4 ↗ all models

Operating Limits

  • Max temperature: 50 °C p.3 ↗
  • Wall clearance back: 100 mm p.4 ↗
  • Wall clearance sidewise: 50 mm p.4 ↗
  • Permitted load: 40 kg p.4 ↗
  • Load per rack: 10 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ CB260 Model series CB, size 260
6-8 p.1 ↗ UL UL variant for 100–120 V / NEMA 5-20 (absent for 200–230 V CEE 7/7 variant)
9-12 p.1 ↗ 120V / 230V Rated voltage variant: 120 V or 230 V
suffix p.1 ↗ -O With O2 control 0.2–20% (absent = standard without O2 control)

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
CB260-230V p.1 ↗ 9040-0207 200...230 V 1~ 50/60 Hz -10,0 A, CEE 7/7, standard
CB260-230V-O p.1 ↗ 9040-0208 200...230 V 1~ 50/60 Hz -10,0 A, CEE 7/7, with O2 control 0.2–20%
CB260UL-120V p.1 ↗ 9040-0209 100...120 V 1~ 50/60 Hz -16,0 A, NEMA 5-20, standard
CB260UL-120V-O p.2 ↗ 9040-0210 100...120 V 1~ 50/60 Hz -16,0 A, NEMA 5-20, with O2 control 0.2–20%