| Temperature range |
0...70 °C
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Temperature range with humidity |
10...70 °C
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Humidity range |
10...80 % RH
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Max. heat compensation at 40°C |
410 W
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Temperature uniformity at 25°C and 60% RH |
0.3 ± K
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Temperature uniformity at 40°C and 75% RH |
0.2 ± K
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Temperature fluctuation at 25°C and 60% RH |
0.1 ± K
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Temperature fluctuation at 40°C and 75% RH |
0.1 ± K
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Humidity fluctuation at 25°C and 60% RH |
1.5 ± % RH
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Humidity fluctuation at 40°C and 75% RH |
1.5 ± % RH
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Rated voltage |
220...240 V
p.2 ↗ |
KBF260-230V |
| Rated voltage |
120 V
p.2 ↗ |
KBF260UL-120V |
| Power frequency |
50/60 Hz
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Nominal power |
1,4 kW
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Unit fuse |
16,0 A
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Phase (nominal voltage) |
1~
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Interior volume |
262 L
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Net weight of the unit (empty) |
188 kg
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Load per rack |
40 kg
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Permitted load |
200 kg
p.2 ↗ |
KBF260-230V / KBF260UL-120V |
| Sound-pressure level at 25°C and 60% RH |
50 dB(A)
p.3 ↗ |
KBF260-230V / KBF260UL-120V |
| Energy consumption at 25°C and 60% RH |
215 Wh/h
p.3 ↗ |
KBF260-230V |
| Energy consumption at 25°C and 60% RH |
255 Wh/h
p.3 ↗ |
KBF260UL-120V |
| Energy consumption at 40°C and 75% RH |
230 Wh/h
p.3 ↗ |
KBF260-230V |
| Energy consumption at 40°C and 75% RH |
295 Wh/h
p.3 ↗ |
KBF260UL-120V |