Sicherheitshinweise MK MKF (E6) 07 2026

Technical Reference Summary

Covered Models & Document Scope

This document (art. no. 7001-0737, issue 07/2026) is a multilingual safety-instructions extract valid only in conjunction with the full operating manuals: MK (E6) art. no. 7001-0714 and MKF (E6) art. no. 7001-0715. It covers the alternating climate chambers with program control of the MK (E6) and MKF (E6) series in sizes 56, 115, 240, 400, 720 and 1020; MKF additionally has a humidity module and CO2 refrigerant. Electrical connection data are given for MK/MKF 240, 400, 720 and 1020 in 50 Hz (400 V) and 60 Hz (480 V) variants with distinct article numbers.

Section-by-Section Technical Breakdown

pp. 8
Chamber overview and interfaces

Front and side views identify the operating panel on the left chamber side, the instrument/signal panels by size, the inspection window, door handle and chamber door. It also locates the refrigeration machine / maintenance access flaps (MK) or refrigeration/humidity module (MKF), main power switch, water-can fill access (MKF), temperature safety device class 2 entry displays, and the communication and signal interfaces.

  • Triangular instrument panel on MK/MKF 56 and 115; triangular signal panel plus instrument panel on MK/MKF 240, 720 and 1020.
  • MK: refrigerating machine and maintenance access flaps; MKF: refrigeration/humidity module with access to fill the water can.
  • Temperature safety device class 2 for over- and under-temperature (option) with entry displays for upper (J1) and lower (J2) temperature limits.
  • Ethernet interface standard; RS485 interface optional.
  • 2 zero-voltage switching outputs via special controller functions (option for MK, regular equipment for MKF).
  • Analog output for temperature (option) and analog output for humidity (option for MKF).
pp. 9, 17
Electrical connection data and installation

The chambers are supplied with a fixed power connection cable of at least 1800 mm. Connection data are tabulated per size and per mains variant (50 Hz 400 V and 60 Hz 480 V) for MK and MKF, including nominal power, nominal current and required over-current release. Installation requires a protective conductor, type B residual current devices and verification of supply voltage against the type plate.

  • Fixed power cable length at least 1800 mm / 70.87 in.
  • Pollution degree 2 per IEC 61010-1; over-voltage category II; housing protection IP 20 per EN 60529; internal over-current release category B, 3-pole.
  • 50 Hz units: 400 V (+/-10%), 3N~; 60 Hz units: 480 V (+/-10%), 3N~.
  • MK 50 Hz: 240 = 5.20 kW / 8.90 A / 3x16 A; 400 = 6.80 kW / 11.90 A / 3x16 A; 720 = 8.70 kW / 14.00 A / 3x16 A; 1020 = 9.60 kW / 22.30 A / 3x32 A.
  • MKF 50 Hz: 240 = 6.80 kW / 10.80 A / 3x16 A; 400 = 7.90 kW / 11.80 A / 3x16 A; 720 = 11.00 kW / 17.80 A / 3x25 A; 1020 = 11.50 kW / 24.70 A / 3x32 A.
  • MK 60 Hz: 240 = 5.20 kW / 7.50 A / 3x16 A; 400 = 6.80 kW / 10.00 A / 3x16 A; 720 = 8.70 kW / 13.50 A / 3x16 A; 1020 = 9.60 kW / 17.60 A / 3x25 A.
  • MKF 60 Hz: 240 = 6.80 kW / 8.50 A / 3x16 A; 400 = 7.90 kW / 10.5 A / 3x16 A; 720 = 11.00 kW / 13.6 A / 3x20 A; 1020 = 11.50 kW / 18.5 A / 3x25 A.
  • Use type B residual current devices and provide sufficient current protection for the number of chambers operated.
  • Check supply voltage against the type plate before connection and start-up; install so the power plug is easily accessible for complete disconnection.
pp. 7, 15
Intended use and MKF water quality

The chambers are intended for conditioning harmless materials, temperature drying and heat treatment of solid, pulverized or bulk material, and material/ageing tests. Solvents must not be explosive or flammable, the operating temperature must be below the flash or sublimation point, and corrosive acids/halides must be excluded. For MKF, suitable fresh-water quality is specified.

  • Indoor use only; installation site and ambient conditions per the operating manual.
  • Not classified as medical devices under Regulation (EU) 2017/745.
  • No explosion protection measures; do not introduce combustible or explosive substances at working temperature, including batteries or lithium-ion batteries.
  • MKF water: deionized water with conductivity from 1 µS/cm up to maximum 20 µS/cm, or water treated by optional BINDER Pure Aqua Service (disposable system).
  • Water in equilibrium with air CO2 with conductivity below 1 µS/cm (ultrapure water) may cause acid corrosion due to low pH.
  • CO2 is used as refrigerant; ensure sufficient room volume calculated per EN 378-1:2016+A1:2020 and provide ventilation or exhaust connection.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Power connection cable is fixed and at least 1800 mm / 70.87 in long. p.9 ↗ Plan the installation position so the supplied cable reaches the socket without extension; the plug must remain easily accessible for emergency disconnection. MK (E6) and MKF (E6) all sizes
Pollution degree 2 per IEC 61010-1, over-voltage category II, housing protection IP 20 per EN 60529, internal over-current release category B 3-pole. p.9 ↗ The chamber is for indoor, dry locations; do not install in damp areas or puddles and protect against splashing water. MK (E6) and MKF (E6) all sizes
50 Hz units require 400 V (+/-10%), 3N~; 60 Hz units require 480 V (+/-10%), 3N~. p.9 ↗ Verify the site supply voltage and frequency against the chamber type plate before connection and start-up. MK (E6) and MKF (E6) all sizes
MK 240 50 Hz: 5.20 kW, 8.90 A, over-current release 3x16 A (art. no. 9x20-0513). p.9 ↗ Size the branch circuit and protection for at least 8.90 A with a 3x16 A release. MK 240, 50 Hz
MK 1020 50 Hz: 9.60 kW, 22.30 A, over-current release 3x32 A (art. no. 9x20-0527). p.9 ↗ The largest MK 50 Hz unit needs a 3x32 A over-current release and a supply rated for at least 22.30 A. MK 1020, 50 Hz
MKF 720 50 Hz: 11.00 kW, 17.80 A, over-current release 3x25 A (art. no. 9x20-0523). p.9 ↗ MKF 720 at 50 Hz requires a 3x25 A release, unlike the MK 720 which uses 3x16 A. MKF 720, 50 Hz
MKF 1020 50 Hz: 11.50 kW, 24.70 A, over-current release 3x32 A (art. no. 9x20-0528). p.9 ↗ Highest 50 Hz current in the range; supply and protection must accommodate 24.70 A with 3x32 A release. MKF 1020, 50 Hz
MK 1020 60 Hz: 9.60 kW, 17.60 A, over-current release 3x25 A (art. no. 9x20-0529). p.9 ↗ At 480 V/60 Hz the MK 1020 needs a smaller release (3x25 A) than at 50 Hz (3x32 A). MK 1020, 60 Hz
MKF 720 60 Hz: 11.00 kW, 13.6 A, over-current release 3x20 A (art. no. 9x20-0525). p.9 ↗ The 60 Hz MKF 720 uses a 3x20 A release, distinct from both the 50 Hz MKF 720 (3x25 A) and the 60 Hz MK 720 (3x16 A). MKF 720, 60 Hz
MKF 1020 60 Hz: 11.50 kW, 18.5 A, over-current release 3x25 A (art. no. 9x20-0530). p.9 ↗ At 480 V/60 Hz the MKF 1020 requires a 3x25 A release and supply rated for at least 18.5 A. MKF 1020, 60 Hz
Use type B residual current devices (RCDs) and provide sufficient current protection according to the number of chambers operated. p.9 ↗ Do not share protection sized for a single chamber when operating multiple units; select type B RCDs. MK (E6) and MKF (E6) all sizes
Maximum switching load of the zero-voltage alarm contacts is 24 V AC/DC, 2.5 A. p.4 ↗ Do not connect devices with higher switching load; exceeding this can damage contacts and the connection socket and creates a shock hazard. MK (E6) and MKF (E6) all sizes
MKF fresh-water supply: deionized water with conductivity from 1 µS/cm up to maximum 20 µS/cm, or water treated by BINDER Pure Aqua Service. p.7 ↗ Water below 1 µS/cm (ultrapure, CO2-equilibrated) may cause acid corrosion due to low pH and must not be used. MKF (E6) all sizes
CO2 is used as refrigerant; required room volume must be calculated per EN 378-1:2016+A1:2020 and a CO2 warning system is recommended. p.7 ↗ Do not set up chambers in non-ventilated recesses; provide technical ventilation or an exhaust connection and sufficient room volume. MKF (E6) all sizes
Operating temperature must lie below the flash point or below the sublimation point of the charging material; no component may release toxic gases. p.7 ↗ Verify the thermal properties of the load before programming a temperature profile. MK (E6) and MKF (E6) all sizes
Loading material must not contain corrosive ingredients such as acids and halides that attack stainless steel, aluminum and copper. p.7 ↗ Exclude acid and halide-containing loads to avoid damage to chamber components. MK (E6) and MKF (E6) all sizes
Temperature safety device class 2 for over- and under-temperature is an option with entry displays for upper (J1) and lower (J2) temperature limits. p.8 ↗ Where fitted, configure both upper and lower limit entries to protect the load and chamber. MK (E6) and MKF (E6) all sizes (option)
2 zero-voltage switching outputs via special controller functions are optional for MK and regular equipment for MKF; analog temperature output is optional and analog humidity output is optional for MKF. p.8 ↗ Specify the option set at ordering if alarm relays or analog signals are required; MKF includes the two switching outputs as standard. MK (E6) and MKF (E6) all sizes

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Humidity system p.6 ↗ Alarm message indicating a fault in the humidity system of MKF chambers; continued operation creates risk of overheating or fire and damage to the chamber and surroundings. Do NOT continue to operate the chamber; do NOT acknowledge the alarm message; turn off the chamber and contact BINDER service. Safety instructions, MKF warning