Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model MK CO₂ 240

Technical Reference Summary

Covered Models & Document Scope

Model MK 240 CO₂ dynamic climate chamber for rapid temperature changes, interior volume 228 L. Two mains variants: MK240CO2-400V (400 V 3~/N/PE 50 Hz, Art.-No. 9020-0513) and MK240CO2-480V (480 V 3~/N/PE 60 Hz, Art.-No. 9020-0515).

Technical Data

Parameter Value Condition
Temperature range -50...180 °C p.3 ↗ MK240CO2-400V, 50 Hz
Temperature uniformity dependent on set 0.1...1.2 ± K p.3 ↗ MK240CO2-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.4 ± K p.3 ↗ MK240CO2-400V, 50 Hz
Cooling down time from 180°C to -40°C 60 min p.3 ↗ MK240CO2-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MK240CO2-400V, 50 Hz
Power frequency 50 Hz p.3 ↗ MK240CO2-400V
Nominal power 5,2 kW p.3 ↗ MK240CO2-400V, 50 Hz
Unit fuse 16,0 A p.3 ↗ MK240CO2-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MK240CO2-400V, 50 Hz
Interior volume 228 L p.3 ↗ MK240CO2-400V, 50 Hz
Net weight of the unit (empty) 311 kg p.3 ↗ MK240CO2-400V, 50 Hz
Load per rack 30 kg p.3 ↗ MK240CO2-400V, 50 Hz
Permitted load 70 kg p.3 ↗ MK240CO2-400V, 50 Hz
Wall clearance back 300 mm p.3 ↗ MK240CO2-400V, 50 Hz
Wall clearance sidewise 200 mm p.3 ↗ MK240CO2-400V, 50 Hz
Width net 1,115 mm p.3 ↗ MK240CO2-400V, 50 Hz
Height net 1,710 mm p.3 ↗ MK240CO2-400V, 50 Hz
Depth net 925 mm p.3 ↗ MK240CO2-400V, 50 Hz
Interior width 735 mm p.4 ↗ MK240CO2-400V, 50 Hz
Interior height 700 mm p.4 ↗ MK240CO2-400V, 50 Hz
Interior depth 443 mm p.4 ↗ MK240CO2-400V, 50 Hz
Unit doors 1 p.4 ↗ MK240CO2-400V, 50 Hz
Sound-pressure level 60 dB(A) p.4 ↗ MK240CO2-400V, 50 Hz
Energy consumption at 25°C 900 Wh/h p.4 ↗ MK240CO2-400V, 50 Hz
Number of shelves (std./max.) 1/6 p.4 ↗ MK240CO2-400V, 50 Hz

Operating Limits

  • Temperature range: -50...180 °C p.3 ↗
  • Wall clearance back: 300 mm p.3 ↗
  • Wall clearance sidewise: 200 mm p.3 ↗
  • Permitted load: 70 kg p.3 ↗
  • Load per rack: 30 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ MK240 Model series MK, size 240
6-8 p.1 ↗ CO2 Refrigerant variant: natural CO₂ (R744)
suffix p.1 ↗ 400V Mains variant 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0513
suffix p.2 ↗ 480V Mains variant 480 V 3~/N/PE 60 Hz, Art.-No. 9020-0515

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
MK240CO2-400V p.1 ↗ 9020-0513 400 V 3~/N/PE 50 Hz, Standard version CEE 16 / 6 H plug 5-pin
MK240CO2-480V p.2 ↗ 9020-0515 480 V 3~/N/PE 60 Hz, Standard version Plug: –

Related models