Data Sheet Model MK 56
Technical Reference Summary
Covered Models & Document Scope
BINDER MK 56 dynamic climate chamber for rapid temperature changes, 60 L interior volume. Two standard variants: MK056-230V (230 V 1~ 50 Hz, CEE 7/7 plug, Art.-No. 9020-0374) and MK056-240V (240 V 1~ 60 Hz, NEMA 6-20 plug, Art.-No. 9020-0388).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -40...180 °C p.2 ↗ | MK056-230V, 50 Hz |
| Temperature uniformity dependent on set value | 0.5...1.5 ± K p.2 ↗ | MK056-230V, 50 Hz |
| Temperature fluctuation dependent on set value | 0.1...0.5 ± K p.2 ↗ | MK056-230V, 50 Hz |
| Cooling down time from 180°C to -40°C | 90 min p.2 ↗ | MK056-230V, 50 Hz |
| Rated voltage | 230 V p.2 ↗ | MK056-230V, 50 Hz |
| Power frequency | 50 Hz p.2 ↗ | MK056-230V |
| Nominal power | 2,8 kW p.2 ↗ | MK056-230V, 50 Hz |
| Unit fuse | 16,0 A p.2 ↗ | MK056-230V, 50 Hz |
| Phase (nominal voltage) | 1~ p.2 ↗ | MK056-230V, 50 Hz |
| Interior volume | 60 L p.2 ↗ | MK056-230V, 50 Hz |
| Net weight of the unit (empty) | 165 kg p.2 ↗ | MK056-230V, 50 Hz |
| Load per rack | 15 kg p.2 ↗ | MK056-230V, 50 Hz |
| Permitted load | 60 kg p.2 ↗ | MK056-230V, 50 Hz |
| Wall clearance back | 300 mm p.2 ↗ | MK056-230V, 50 Hz |
| Wall clearance sidewise | 200 mm p.2 ↗ | MK056-230V, 50 Hz |
| Width net | 720 mm p.2 ↗ | MK056-230V, 50 Hz |
| Height net | 1,445 mm p.2 ↗ | MK056-230V, 50 Hz |
| Depth net | 778 mm p.2 ↗ | MK056-230V, 50 Hz |
| Interior width | 400 mm p.3 ↗ | MK056-230V, 50 Hz |
| Interior height | 420 mm p.3 ↗ | MK056-230V, 50 Hz |
| Interior depth | 350 mm p.3 ↗ | MK056-230V, 50 Hz |
| Unit doors | 1 p.3 ↗ | MK056-230V, 50 Hz |
| Sound-pressure level | 59 dB(A) p.3 ↗ | MK056-230V, 50 Hz |
| Energy consumption at 20°C | 600 Wh/h p.3 ↗ | MK056-230V, 50 Hz |
| Number of shelves (std./max.) | 1/4 p.3 ↗ | MK056-230V, 50 Hz |
Operating Limits
- Temperature range: -40...180 °C p.2 ↗
- Permitted load: 60 kg p.2 ↗
- Load per rack: 15 kg p.2 ↗
- Wall clearance back: 300 mm p.2 ↗
- Wall clearance sidewise: 200 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-5 p.1 ↗ | MK056 | Model series MK (dynamic climate chamber) and size 56 |
| 7-9 p.1 ↗ | 230 / 240 | Rated voltage variant: 230 V or 240 V |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| MK056-230V p.1 ↗ | 9020-0374 | 230 V 1~ 50 Hz, CEE 7/7 plug, Standard | Standard model version | |
| MK056-240V p.1 ↗ | 9020-0388 | 240 V 1~ 60 Hz, NEMA 6-20 plug, Standard | Standard model version |