Data Sheet Model VD 23
Technical Reference Summary
Covered Models & Document Scope
Model VD 23 vacuum drying chamber for non-flammable solvents, interior volume 24 L. Two standard variants: VD023-230V (200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9630-0001) and VD023UL-120V (100...120 V 1~ 50/60 Hz, NEMA 5-15 plug, Art.-No. 9630-0005).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +9 °C above ambient temperature to 220 °C p.2 ↗ | VD023-230V |
| Temperature uniformity at 100°C | 1.0 ± K p.2 ↗ | VD023-230V |
| Temperature fluctuation at 100°C | 0.1 ± K p.2 ↗ | VD023-230V |
| Heating up time to 100°C | 80 min p.2 ↗ | VD023-230V |
| Leak rate | 0,01 bar/h p.2 ↗ | VD023-230V |
| Pressure fluctuation over time at 50 mbar | 1 ± mbar p.2 ↗ | VD023-230V |
| Pressure fluctuation over time at 400 mbar | 70 ± mbar p.2 ↗ | VD023-230V |
| Rated voltage | 200...230 V p.2 ↗ | VD023-230V |
| Rated voltage | 100...120 V p.2 ↗ | VD023UL-120V |
| Power frequency | 50/60 Hz p.2 ↗ | VD023-230V |
| Nominal power | 0,9 kW p.2 ↗ | VD023-230V |
| Unit fuse | 6,3 A p.2 ↗ | VD023-230V |
| Unit fuse | 10,0 A p.2 ↗ | VD023UL-120V |
| Phase (nominal voltage) | 1~ p.2 ↗ | VD023-230V |
| Vacuum connection with small flange | 16 DN mm p.2 ↗ | VD023-230V |
| Measuring access port with small flange | 16 DN mm p.2 ↗ | VD023-230V |
| Interior volume | 24 L p.2 ↗ | VD023-230V |
| Net weight of the unit (empty) | 64 kg p.2 ↗ | VD023-230V |
| Load per rack | 20 kg p.2 ↗ | VD023-230V |
| Permitted load | 50 kg p.2 ↗ | VD023-230V |
| Wall clearance back | 100 mm p.2 ↗ | VD023-230V |
| Wall clearance sidewise | 70 mm p.2 ↗ | VD023-230V |
| Width net | 523 mm p.3 ↗ | VD023-230V |
| Height net | 698 mm p.3 ↗ | VD023-230V |
| Depth net | 413 mm p.3 ↗ | VD023-230V |
Operating Limits
- Temperature range: +9 °C above ambient temperature to +220 °C p.2 ↗
- Permitted load: 50 kg p.2 ↗
- Load per rack: 20 kg p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
- Wall clearance sidewise: 70 mm p.2 ↗
- Number of shelves (std./max.): 1/4 p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-5 p.1 ↗ | VD023 | Model series VD (vacuum drying chamber), size 23 |
| suffix p.1 ↗ | UL | UL variant with 100...120 V power supply and NEMA 5-15 plug (absent = standard 200...230 V, CEE 7/7) |
| suffix p.1 ↗ | 230V | Mains voltage variant 200...230 V 1~ 50/60 Hz |
| suffix p.1 ↗ | 120V | Mains voltage variant 100...120 V 1~ 50/60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| VD023-230V p.1 ↗ | 9630-0001 | 200...230 V 1~ 50/60 Hz, 6,3 A, CEE 7/7 plug, Standard | Standard model version | |
| VD023UL-120V p.1 ↗ | 9630-0005 | 100...120 V 1~ 50/60 Hz, 10,0 A, NEMA 5-15 plug, Standard | UL model version |