Data Sheet Model VD 56

Technical Reference Summary

Covered Models & Document Scope

BINDER VD 56 vacuum drying chamber for non-flammable solvents, interior volume 55 L. Two standard variants: VD056-230V (200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9630-0002) and VD056UL-120V (100...120 V 1~ 50/60 Hz, NEMA 5-15 plug, Art.-No. 9630-0006).

Technical Data

Parameter Value Condition
Interior volume 55 L p.2 ↗ VD 56
Temperature range +9 °C above ambient temperature to 220 °C p.2 ↗ VD 56
Temperature uniformity at 100°C 1.2 ± K p.2 ↗ VD 56
Temperature fluctuation at 100°C 0.1 ± K p.2 ↗ VD 56
Heating up time to 100°C 80 min p.2 ↗ VD 56
Leak rate 0,01 bar/h p.2 ↗ VD 56
Pressure fluctuation over time at 50 mbar 1 ± mbar p.2 ↗ VD 56
Pressure fluctuation over time at 400 mbar 70 ± mbar p.2 ↗ VD 56
Rated voltage 200...230 V p.2 ↗ VD056-230V
Rated voltage 100...120 V p.2 ↗ VD056UL-120V
Power frequency 50/60 Hz p.2 ↗ VD 56
Nominal power 1,4 kW p.2 ↗ VD 56
Unit fuse 8,0 A p.2 ↗ VD056-230V
Unit fuse 16,0 A p.2 ↗ VD056UL-120V
Phase (nominal voltage) 1~ p.2 ↗ VD 56
Vacuum connection with small flange 16 DN mm p.2 ↗ VD 56
Measuring access port with small flange 16 DN mm p.2 ↗ VD 56
Net weight of the unit (empty) 96 kg p.2 ↗ VD 56
Load per rack 20 kg p.2 ↗ VD 56
Permitted load 60 kg p.2 ↗ VD 56
Wall clearance back 100 mm p.2 ↗ VD 56
Wall clearance sidewise 70 mm p.2 ↗ VD 56
Width net 638 mm p.3 ↗ VD 56
Height net 815 mm p.3 ↗ VD 56
Depth net 461 mm p.3 ↗ VD 56

Operating Limits

  • Temperature range: +9 °C above ambient temperature to +220 °C p.2 ↗
  • Permitted load: 60 kg p.2 ↗
  • Load per rack: 20 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 70 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ VD056 Model series VD (vacuum drying chamber) and size 56
suffix p.1 ↗ UL UL variant with 100...120 V power supply and NEMA 5-15 plug (absent = standard 200...230 V CEE 7/7)
suffix p.1 ↗ 230V Standard variant rated voltage 200...230 V
suffix p.1 ↗ 120V UL variant rated voltage 100...120 V

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
VD056-230V p.1 ↗ 9630-0002 200...230 V 1~ 50/60 Hz, 8,0 A, CEE 7/7 plug, Standard Standard model version
VD056UL-120V p.1 ↗ 9630-0006 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-15 plug, Standard UL model version